COVID-19驾车穿越检测中鼻咽拭子铝柄断裂:影像学检出滞留异物

Fractured aluminum nasopharyngeal swab during drive-through testing for COVID-19: radiographic detection of a retained foreign body.
PubMed 文献 · Skeletal Radiol · 2020-08-21

本文报道一例在驾车穿越检测点采集鼻咽拭子时铝柄断裂、疑似滞留于无症状患者体内的病例,初始覆盖颅底、颈部、胸部及腹部的系列X线摄影未显示异物,后经内镜发现并取出。作者利用体模模型探索出一种低管电压、紧准直并以鼻骨区为中心的X线摄影技术,能最佳显示铝柄。作者认为铝制异物X线摄影可能难以显示,优化摄影技术或有助于临床处理,对疑似而X线阴性的病例可考虑进一步行CT或内镜检查。

要点

  • 鼻咽拭子铝柄断裂滞留时,常规覆盖颅底、颈部、胸腹部的X线摄影可能无法显示异物。
  • 采用低管电压、紧准直并以鼻骨区为中心的X线摄影技术,在体模中可最佳显示铝柄。
  • 对疑似异物滞留而X线阴性的病例,作者认为可考虑进一步行CT或内镜检查。

核心结论

  • 铝制异物可能难以在X线摄影上显示,优化X线摄影技术可能有助于指导不寻常病例的临床处理。
  • 对疑似病例而X线阴性时,作者认为可考虑进一步行计算机断层扫描或内镜检查。

研究设计与样本

  • 病例报告:一例无症状患者在驾车穿越检测点采集鼻咽拭子时铝柄断裂并疑似滞留。
  • 初始评估采用覆盖颅底、颈部、胸部及腹部的系列X线摄影。
  • 采用体模模型探索能更好显示鼻咽拭子铝柄的X线摄影技术。

关键影像表现/征象

  • 初始系列X线摄影(覆盖颅底、颈部、胸部及腹部)未显示断裂的拭子。
  • 进一步临床评估经内镜发现拭子嵌顿于左侧下鼻甲与鼻底之间,由耳鼻喉科医生取出。
  • 体模模型中,采用较低管电压(kVp)并紧准直、以鼻骨区为中心的技术对铝柄显示最佳。

临床应用与阅片要点

  • 识别铝制异物在X线摄影上可能难以显示,并优化X线摄影采集技术,可能有助于指导不寻常病例的临床处理。
  • 对疑似病例而X线摄影阴性时,作者认为可考虑进一步行计算机断层扫描或内镜检查。
  • 本研究报告的优化技术为较低管电压(kVp)并紧准直、以鼻骨区为中心。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

The ongoing coronavirus disease 2019 (COVID-19) pandemic has increased the need for safe and efficient testing as a key containment strategy. Drive-through testing with nasopharyngeal swab has been implemented in many places in the USA as it allows for expeditious testing of large numbers of patients, limits healthcare workers' risk of exposure, and minimizes the use of personal protective equipment. We present a case where the aluminum shaft of the nasopharyngeal swab fractured during specimen collection at a drive-through testing facility and was suspected to have remained in the asymptomatic patient. Initial evaluation with a series of radiographs covering the skull base, neck, chest, and abdomen did not reveal the swab. On further clinical evaluation, the swab was found endoscopically, lodged between the left inferior turbinate and nasal floor, and was removed by an otorhinolaryngologist. Using a phantom model, we aimed to delineate an imaging technique to better visualize the aluminum shaft of the nasopharyngeal swab on radiographs to help in identification. A technique using lower tube voltage (kVp) with tight collimation centered at the nasal bones area produced the best visualization of the aluminum shaft of the swab. Recognition that aluminum foreign bodies may be difficult to visualize radiographically and optimization of radiograph acquisition technique may help guide clinical management in unusual cases. Further evaluation with computed tomography or endoscopy should be considered in suspected cases where radiographs are negative.