文献速递神经介入CTDSA

虚拟稀释锥形束CT技术(Vessel ASSIST)可视化颅内动脉瘤治疗中血流导向支架的壁贴附

Visualization of flow diverter stent wall apposition during intracranial aneurysm treatment using a virtually diluted cone beam CT technique (Vessel ASSIST).
PubMed 文献 · Neuroradiology · 2020-08-01

该研究回顾性评估了27例颅内动脉瘤血流导向支架治疗患者,比较虚拟稀释(VD)软件处理的双锥形束CT图像与常规二维DSA对支架释放和壁贴附的评估。VD提高了诊断信心(评分P<0.001)和读者间一致性(P<0.0001),X线曝光无显著差异。

要点

  • VD技术可提高FD壁贴附评估的诊断信心,评分达4.9±0.2。
  • VD使读者间一致性从中等(0.49324)提升至良好(0.7272)。
  • VD与常规2D DSA的X线曝光无显著差异(剂量面积乘积P=0.54)。

诊断标准与关键数值

  • FD释放和壁贴附评分:DSA为4.2±1.0(R1)和4.0±1.1(R2),VD为3.7±1.2(R1)和4.1±1.0(R2)。
  • 诊断信心评分:DSA为3.7±0.7(R1)和4.0±0.7(R2),VD为4.9±0.2(R1)和4.9±0.2(R2),P<0.001。
  • 读者间一致性:2D DSA为中等(0.49324),VD提升至良好(0.7272),P<0.0001。
  • 释放评估的读者间一致性(P=0.68)和剂量面积乘积(P=0.54)在两技术间无显著差异。

临床应用与阅片要点

  • VD成像可用于FD释放后壁贴附的评估,提高诊断信心和读者间一致性。
  • VD与常规2D DSA的X线曝光相当,未增加辐射剂量。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: Flow diverters (FD) have poor radiopacity, challenging visualization of deployment and vessel wall apposition with conventional neuroimaging modalities. We evaluated a novel cone beam computed tomography (CT) imaging technique that allows virtual dilution (VD) of contrast media to facilitate workflow and ensure accurate assessment of FD wall apposition.

METHODS: We retrospectively evaluated all patients treated for intracranial aneurysms with FD at our institution between November 2018 and November 2019. Undiluted injected dual cone beam CT acquisitions performed post-stenting were displayed with VD software (GE Healthcare). The resulting images were compared with conventional two-dimensional (2D) digital subtraction angiography (DSA) images. Two neurointerventionalists (Reader 1 and Reader 2, (R1, R2)) independently assessed FD deployment and wall apposition. Confidence in the diagnosis, inter-reader agreement, and X-ray exposure were assessed.

RESULTS: A total of 27 cases were reviewed. FD deployment and wall apposition scores were 4.2 ± 1.0 (R1) and 4.0 ± 1.1 (R2) for DSA and 3.7 ± 1.2 (R1) and 4.1 ± 1.0 (R2) for VD. Confidence in the diagnosis was improved with VD, with scores of 3.7 ± 0.7 (R1) and 4.0 ± 0.7 (R2) using DSA and 4.9 ± 0.2 (R1) and 4.9 ± 0.2 (R2) using VD (P < 0.001). Inter-reader agreement using 2D DSA was improved from moderate (0.49324) to good (0.7272) (P < 0.0001). There were no significant differences in inter-reader agreement in the deployment assessment (P = 0.68) or dose-area product (P = 0.54) between techniques.

CONCLUSION: VD imaging with dual cone beam CT enables accurate assessment of FD wall apposition after deployment with greater confidence and improved inter-reader agreement versus conventional 2D DSA alone, with comparable X-ray exposure.

来源:Neuroradiology,查看原文。版权归原作者所有。

引用

  1. Hassan AE, Wise J, Burke EM, et al. Visualization of flow diverter stent wall apposition during intracranial aneurysm treatment using a virtually diluted cone beam CT technique (Vessel ASSIST). Neuroradiology. 2020. PMID: 32740710. doi:10.1007/s00234-020-02507-8

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。