Woven EndoBridge治疗颅内动脉瘤后形状改变常见:一项纵向定量分析研究
本研究对30例患者32个WEB治疗的颅内动脉瘤,在基线及随访CT上定量分析WEB形状改变与动脉瘤闭塞的关系。中位随访11.4个月,90.6%的WEB装置高度减少至少5%,1年中位高度减少19.2%,且高度减少随时间间隔延长而加重(P<.0001);动脉瘤闭塞质量与WEB形状改变无显著关联。
要点
- WEB治疗颅内动脉瘤后形状改变常见,90.6%装置高度减少至少5%。
- WEB高度减少随时间延长而加重,1年中位减少19.2%(范围8.6%-52.3%)。
- WEB形状改变与动脉瘤闭塞质量、装置尺寸、破裂状态、位置及再干预率无显著关联。
核心结论
- WEB治疗的颅内动脉瘤中形状改变常见,1年中位高度减少19.2%;动脉瘤闭塞质量与WEB形状改变无显著关联。
研究设计与样本
- 纵向定量分析研究,纳入30例患者32个WEB治疗的颅内动脉瘤,满足基线(介入当日)及至少>1个月后随访头颅CT的纳入标准,中位随访时间11.4个月(四分位距6.5-21.5个月),WEB形状改变通过CT扫描上的半自动方法定量,采用线性混合效应回归模型进行关联分析。
关键影像表现/征象
- 29/32个WEB装置(90.6%)高度减少至少5%。
- WEB高度减少随时间间隔(天)延长而显著加重(P<.0001),1年中位减少19.2%(范围8.6%-52.3%)。
- 充分闭塞率为84.4%,动脉瘤残留率为15.6%。
诊断标准与关键数值
- WEB形状改变定义为CT上高度减少至少5%。
- 充分闭塞率84.4%,动脉瘤残留率15.6%。
临床应用与阅片要点
- 随访CT上可定量评估WEB装置高度减少,注意其随时间延长而加重的趋势。
- WEB形状改变与动脉瘤闭塞质量无显著关联,阅片时需结合闭塞情况综合判断。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: After endovascular treatment of intracranial aneurysms with the Woven EndoBridge (WEB) device, worsening of aneurysm occlusion or re-opening was reported to be possibly associated with WEB shape modification. This study analyzed quantitatively the WEB shape modification with time in association with anatomic results.
MATERIALS AND METHODS: Thirty patients with 32 WEB-treated intracranial aneurysms fulfilled the inclusion criteria of cranial CT at baseline (day of intervention) and a follow-up CT at least >1 months' later (median follow-up time, 11.4 months; interquartile range, 6.5-21.5 months). Adequate occlusion was observed in 84.4%, and aneurysm remnant, in 15.6%. WEB shape modification was quantified by a semiautomated approach on CT scans. Time courses were evaluated graphically and analytically; association analyses were performed by linear mixed-effects regression models.
RESULTS: In 29/32 WEB devices (90.6%), the reduction in height was found to be at least 5%. The decrease in height with time was significantly associated with the time interval in days since the intervention (P < .0001): A stronger decrease in WEB height was associated with a longer time since the intervention (median reduction in 1 year, 19.2%; range, 8.6%-52.3%). No significant association was found with the quality of aneurysm occlusion, device size, rupture status of the aneurysm, aneurysm location, and reintervention rate.
CONCLUSIONS: Shape modification was common in WEB-treated intracranial aneurysms with a median height reduction of 19.2% in 1 year. The quality of aneurysm occlusion was not associated with WEB modification.
引用
- Rosskopf J, Braun M, Dreyhaupt J, et al. Shape Modification is Common in Woven EndoBridge-Treated Intracranial Aneurysms: A Longitudinal Quantitative Analysis Study. AJNR Am J Neuroradiol. 2020. PMID: 32732267. doi:10.3174/ajnr.A6669
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。