自动三维超声成像技术用于胎儿头部生物测量的准确性
这项前瞻性观察研究纳入143例孕18+0至22+6周行常规解剖筛查的孕妇,比较操作者手动与基于人工智能工具SonoCNS® Fetal Brain的自动三维超声测量胎儿头部结构的准确性。自动测量双顶径和头围的组内相关系数为0.80-0.88,信度良好;经小脑直径、小脑延髓池和侧脑室后角的组内相关系数为0.23-0.50,信度差至中等。作者认为该技术可靠识别和测量双顶径与头围,但在纳入常规超声方案前仍需进一步优化。
要点
- 自动三维超声测量双顶径和头围与手动测量信度良好,组内相关系数0.80-0.88。
- 自动测量经小脑直径、小脑延髓池和侧脑室后角信度差至中等,组内相关系数0.23-0.50。
- 操作者首次尝试即成功获取全部自动测量的比例分别为超声技师70%、医师76%。
核心结论
- 自动三维超声成像使用SonoCNS可可靠识别和测量双顶径与头围,但在准确识别和测量经小脑直径、小脑延髓池和侧脑室后角方面一致性较差。
研究设计与样本
- 前瞻性观察研究,纳入143例孕18+0至22+6周行常规解剖筛查的患者,每例由超声技师和医师各连续进行一次胎儿头部超声检查,先手动测量双顶径、头围、经小脑直径、小脑延髓池和侧脑室后角,再使用基于人工智能的工具SonoCNS® Fetal Brain进行自动测量。
诊断标准与关键数值
- 自动测量与手动测量比较的组内相关系数:双顶径和头围为0.80-0.88(信度良好);经小脑直径、小脑延髓池和侧脑室后角为0.23-0.50(信度差至中等)。
临床应用与阅片要点
- 自动三维超声测量双顶径和头围与手动测量信度良好,但经小脑直径、小脑延髓池和侧脑室后角的自动测量信度差至中等,作者认为在将该技术纳入常规超声方案前需进一步优化。
局限性
- 作者结论指出,虽然结果令人鼓舞,但在将该技术纳入常规超声方案前,需对自动技术进一步优化。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To evaluate the accuracy of an automated three-dimensional (3D) ultrasound technique for fetal intracranial measurements compared with manual acquisition.
METHODS: This was a prospective observational study of patients presenting for routine anatomical survey between 18 + 0 and 22 + 6 weeks' gestation. After providing informed consent, each patient underwent two consecutive ultrasound examinations of the fetal head, one by a sonographer and one by a physician. Each operator obtained manual measurements of the biparietal diameter (BPD), head circumference (HC), transcerebellar diameter (TCD), cisterna magna (CM) and posterior horn of the lateral ventricle (Vp), followed by automated measurements of these structures using an artificial intelligence-based tool, SonoCNS® Fetal Brain. Both operators repeated the automated approach until all five measurements were obtained in a single sweep, up to a maximum of three attempts. The accuracy of automated measurements was compared with that of manual measurements using intraclass correlation coefficients (ICC) by operator type, accounting for patient and ultrasound characteristics.
RESULTS: One hundred and forty-three women were enrolled in the study. Median body mass index was 24.0 kg/m2 (interquartile range (IQR), 22.5-26.8 kg/m2 ) and median subcutaneous thickness was 1.6 cm (IQR, 1.3-2.0 cm). Fifteen (10%) patients had at least one prior Cesarean delivery, 17 (12%) had other abdominal surgery and 78 (55%) had an anterior placenta. Successful acquisition of the automated measurements was achieved on the first, second and third attempts in 70%, 22% and 3% of patients, respectively, by sonographers and in 76%, 16% and 3% of cases, respectively, by physicians. The automated algorithm was not able to identify and measure all five structures correctly in six (4%) and seven (5%) patients scanned by the sonographers and physicians, respectively. The ICCs reflected good reliability (0.80-0.88) of the automated compared with the manual approach for BPD and HC and poor to moderate reliability (0.23-0.50) for TCD, CM and Vp. Fetal lie, head position, placental location, maternal subcutaneous thickness and prior Cesarean section were not associated with the success or accuracy of the automated technique.
CONCLUSIONS: Automated 3D ultrasound imaging of the fetal head using SonoCNS reliably identified and measured BPD and HC but was less consistent in accurately identifying and measuring TCD, CM and Vp. While these results are encouraging, further optimization of the automated technology is necessary prior to incorporation of the technique into routine sonographic protocols. © 2020 International Society of Ultrasound in Obstetrics and Gynecology.
引用
- Pluym ID, Afshar Y, Holliman K, et al. Accuracy of automated three-dimensional ultrasound imaging technique for fetal head biometry. Ultrasound Obstet Gynecol. 2021. PMID: 32770786. doi:10.1002/uog.22171
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。