文献速递外周血管介入胸部DSACT

DynaCT血管造影重建在球囊肺动脉成形术中的应用

Application of DynaCT angiographic reconstruction in balloon pulmonary angioplasty.
PubMed 文献 · Eur Radiol · 2020-07-03

该研究纳入23例CTEPH患者的34次球囊肺动脉成形术(BPA),比较DynaCT血管造影重建引导(3D组,23次)与DSA二维血管造影引导(2D组,11次)的可行性与准确性。3D组单次BPA处理的靶血管数更多(5.83±2.33 vs 3.73±1.10,p=0.008)、手术时间更短(3.58±0.61 vs 4.49±0.91 h,p=0.002)、剂量面积乘积(DAP)更低(4682.82±1950.64 vs 13901.82±5549.69,p<0.001)、对比剂用量更少(225.22±48.70 vs 292.73±76.82 mL,p=0.013)。作者结论认为DynaCT血管造影重建引导BPA可行且准确,可能有助于优化手术流程并减少辐射暴露。

要点

  • DynaCT血管造影重建引导BPA可行且准确,单次可处理更多靶血管(5.83±2.33 vs 3.73±1.10)。
  • 3D组手术时间更短(3.58±0.61 vs 4.49±0.91 h),对比剂用量更少(225.22±48.70 vs 292.73±76.82 mL)。
  • 3D组DAP显著低于2D组(4682.82±1950.64 vs 13901.82±5549.69,p<0.001)。

诊断标准与关键数值

  • 3D组单次BPA处理靶血管数为5.83±2.33条,2D组为3.73±1.10条(p=0.008)。
  • 3D组手术时间为3.58±0.61 h,2D组为4.49±0.91 h(p=0.002)。
  • 3D组DAP为4682.82±1950.64,2D组为13901.82±5549.69(p<0.001)。
  • 3D组对比剂用量为225.22±48.70 mL,2D组为292.73±76.82 mL(p=0.013)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To investigate the feasibility and accuracy of balloon pulmonary angioplasty (BPA) using DynaCT angiographic reconstruction guidance.

METHODS: Thirty-four BPAs (23 CTEPH patients) targeting 175 pulmonary arteries were included. Eleven BPAs (2D group) were guided by DSA two-dimensional angiography. Another twenty-three BPAs (3D group) were guided using DynaCT angiographic reconstruction. The volume rendering (VR) method was used to obtain a three-dimensional image of the blood vessels. This image was used as a reference to continue BPA treatment under the guidance of vascular three-dimensional reconstruction technology. Procedure durations and radiation exposure data were compared between the two groups using Mann-Whitney U test.

RESULTS: Using the DynaCT angiographic reconstruction technique, more target vessels were treated in a single BPA procedure (5.83 ± 2.33 vs 3.73 ± 1.10 vessels per BPA, p = 0.008) in a shorter operation time (3.58 ± 0.61 vs 4.49 ± 0.91 h, p = 0.002). Overall, the dose area product (DAP) was significantly higher for the 2D group than for the 3D group (13,901.82 ± 5549.69 vs 4682.82 ± 1950.64, p < 0.001). The use of the DynaCT angiographic reconstruction technique to guide BPA required a lower dose of contrast agent (225.22 ± 48.70 vs 292.73 ± 76.82 mL, p = 0.013) and less radiation exposure.

CONCLUSIONS: The use of DynaCT angiographic reconstruction guidance in patients undergoing BPA is feasible and accurate. Images of DynaCT angiographic reconstruction may be beneficial for optimizing the operative process in BPA with reduced radiation exposure.

KEY POINTS: • BPA guidance by DynaCT angiographic reconstruction is feasible and accurate. • DynaCT angiographic reconstruction may be beneficial for optimizing the operative process. • DynaCT angiographic reconstruction can reduce patient radiation dose due to multi-times of BPA sessions.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Lin JL, Chen HM, Lin FC, et al. Application of DynaCT angiographic reconstruction in balloon pulmonary angioplasty. Eur Radiol. 2020. PMID: 32621239. doi:10.1007/s00330-020-07028-9

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。