迭代金属伪影减少与双能CT在肩关节置换假体金属伪影抑制中的获益

Benefits of iterative metal artifact reduction and dual-energy CT towards mitigating artifact in the setting of total shoulder prostheses.
PubMed 文献 · Skeletal Radiol · 2020-06-29

纳入6例男性和13例女性肩关节置换术后患者,行双能肩部CT并重建四组图像,由3名读片者盲法评估伪影程度与图像质量。结果显示130 keV双能虚拟单能联合迭代金属伪影减少组的伪影评分、整体图像质量评分及CT值差值均优于原始多能kV图像。

要点

  • 130 keV双能虚拟单能联合迭代金属伪影减少可改善骨与软组织显示。
  • 联合方案伪影评分3.02优于原始多能kV图像的4.28(P<0.001)。
  • 联合方案整体图像质量中位评分3.00优于原始多能kV图像的4.33(P<0.001)。

诊断标准与关键数值

  • 130 keV双能虚拟单能联合迭代金属伪影减少组平均伪影评分3.02,优于原始多能kV图像的4.28(P<0.001)。
  • 联合方案组整体图像质量中位评分3.00,优于原始多能kV图像的4.33(P<0.001)。
  • 联合方案组受伪影影响区域与正常区域CT值中位差值:骨区72.28,低于原始多能kV图像的252.08(P<0.001)。
  • 联合方案组软组织区CT值中位差值15.09,低于原始多能kV图像的324.38(P<0.001)。

临床应用与阅片要点

  • 对肩关节置换术后金属伪影明显的CT,可参考采用130 keV双能虚拟单能联合迭代金属伪影减少重建以改善骨与软组织显示。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To determine the utility of iterative metal artifact reduction and 130 keV dual-energy virtual monoenergetic images to improve bone and soft tissue visualization in CT scans affected by metal artifacts.

MATERIAL AND METHODS: Thirteen females and 6 males with a history of total shoulder prosthesis who underwent dual-energy shoulder CT were included. Four sets of images were reconstructed for each patient: (1) original polychromatic kV images reconstructed with weighted filtered back projection; (2) polychromatic kV images with iterative metal artifact reduction; (3) 130 keV dual-energy virtual monoenergetic; (4) combined iterative metal artifact reduction and 130 keV dual-energy virtual monoenergetic. Three readers blindly reviewed all image sets and graded the extent of artifact and image quality.

RESULTS: Mean artifact score and median overall image quality score were better in 130 keV dual-energy virtual monoenergetic with iterative metal artifact reduction compared with those in original polychromatic kV images (3.02 vs 4.28, P < 0.001 and 3.00 vs 4.33, P < 0.001, respectively). The median difference in CT numbers between regions affected by artifacts and normal regions was lowest in 130 keV dual-energy virtual monoenergetic with iterative metal artifact reduction compared with that in original polychromatic kV images (72.28 vs 252.08, P < 0.001 for bony regions and 15.09 vs 324.38, P < 0.001 for soft tissue).

CONCLUSION: In patients with metal artifacts due to shoulder replacement surgery, the use of dual-energy monoenergetic images and iterative metal artifact reduction reconstruction significantly improves both subjective and objective indicators of image quality.