迭代金属伪影减少与双能CT在肩关节置换假体金属伪影抑制中的获益
纳入6例男性和13例女性肩关节置换术后患者,行双能肩部CT并重建四组图像,由3名读片者盲法评估伪影程度与图像质量。结果显示130 keV双能虚拟单能联合迭代金属伪影减少组的伪影评分、整体图像质量评分及CT值差值均优于原始多能kV图像。
要点
- 130 keV双能虚拟单能联合迭代金属伪影减少可改善骨与软组织显示。
- 联合方案伪影评分3.02优于原始多能kV图像的4.28(P<0.001)。
- 联合方案整体图像质量中位评分3.00优于原始多能kV图像的4.33(P<0.001)。
诊断标准与关键数值
- 130 keV双能虚拟单能联合迭代金属伪影减少组平均伪影评分3.02,优于原始多能kV图像的4.28(P<0.001)。
- 联合方案组整体图像质量中位评分3.00,优于原始多能kV图像的4.33(P<0.001)。
- 联合方案组受伪影影响区域与正常区域CT值中位差值:骨区72.28,低于原始多能kV图像的252.08(P<0.001)。
- 联合方案组软组织区CT值中位差值15.09,低于原始多能kV图像的324.38(P<0.001)。
临床应用与阅片要点
- 对肩关节置换术后金属伪影明显的CT,可参考采用130 keV双能虚拟单能联合迭代金属伪影减少重建以改善骨与软组织显示。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To determine the utility of iterative metal artifact reduction and 130 keV dual-energy virtual monoenergetic images to improve bone and soft tissue visualization in CT scans affected by metal artifacts.
MATERIAL AND METHODS: Thirteen females and 6 males with a history of total shoulder prosthesis who underwent dual-energy shoulder CT were included. Four sets of images were reconstructed for each patient: (1) original polychromatic kV images reconstructed with weighted filtered back projection; (2) polychromatic kV images with iterative metal artifact reduction; (3) 130 keV dual-energy virtual monoenergetic; (4) combined iterative metal artifact reduction and 130 keV dual-energy virtual monoenergetic. Three readers blindly reviewed all image sets and graded the extent of artifact and image quality.
RESULTS: Mean artifact score and median overall image quality score were better in 130 keV dual-energy virtual monoenergetic with iterative metal artifact reduction compared with those in original polychromatic kV images (3.02 vs 4.28, P < 0.001 and 3.00 vs 4.33, P < 0.001, respectively). The median difference in CT numbers between regions affected by artifacts and normal regions was lowest in 130 keV dual-energy virtual monoenergetic with iterative metal artifact reduction compared with that in original polychromatic kV images (72.28 vs 252.08, P < 0.001 for bony regions and 15.09 vs 324.38, P < 0.001 for soft tissue).
CONCLUSION: In patients with metal artifacts due to shoulder replacement surgery, the use of dual-energy monoenergetic images and iterative metal artifact reduction reconstruction significantly improves both subjective and objective indicators of image quality.
引用
- Mohammadinejad P, Baffour FI, Adkins MC, et al. Benefits of iterative metal artifact reduction and dual-energy CT towards mitigating artifact in the setting of total shoulder prostheses. Skeletal Radiol. 2020. PMID: 32601733. doi:10.1007/s00256-020-03528-3
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。