超短回波时间(UTE)成像揭示老年人结合水变化对亚临床肌腱病敏感
研究采用超短回波时间磁共振成像定量13名年轻和13名老年人跟腱的结合水与自由水横向弛豫时间及结合水分数,并由肌骨放射科医生对解剖图像分级识别亚临床肌腱病征象。老年人跟腱结合水横向弛豫时间延长60%、结合水分数减小5%;其中7条有亚临床肌腱病征象的跟腱结合水分数较正常跟腱减小7%且结合水横向弛豫时间显著延长。
要点
- 老年人跟腱结合水横向弛豫时间较年轻人延长60%,结合水分数减小5%。
- 解剖图像显示跟腱增厚和信号增高提示亚临床肌腱病。
- 有亚临床肌腱病征象的跟腱结合水分数减小7%且结合水横向弛豫时间显著延长。
关键影像表现/征象
- 老年人跟腱结合水横向弛豫时间较年轻人延长60%(p=0.004)。
- 老年人跟腱自由水横向弛豫时间与年轻人相似(p=0.86)。
- 老年人跟腱结合水分数较年轻人减小5%(p=0.048)。
- 7条老年人跟腱出现跟腱增厚和信号强度增高,提示亚临床肌腱病。
- 有亚临床肌腱病征象的跟腱结合水分数较年轻人和老年人的正常跟腱减小7%(p=0.02),结合水横向弛豫时间显著延长(p<0.001)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Use ultrashort echo time (UTE) magnetic resonance imaging to quantify bound water components of asymptomatic older Achilles tendons and investigate the relationship between UTE findings and imaging assessment of sub-clinical tendinopathy.
MATERIALS AND METHODS: Thirteen young (age 25 ± 4.8) and thirteen older (age 67 ± 4.7) adults were tested. A UTE sequence was used to quantify the transverse relaxation times of bound ([Formula: see text]) and free ([Formula: see text]) water and the bound water fraction (Fs) in the Achilles tendon. Anatomical images were collected and graded by a musculoskeletal radiologist to identify signs of sub-clinical tendinopathy. Two-sample t tests were used to compare [Formula: see text], [Formula: see text], and Fs between age groups and between adults with and without sub-clinical tendinopathy.
RESULTS: Older tendons exhibited a 60% longer [Formula: see text] (p = 0.004), similar [Formula: see text] (p = 0.86), and 5% smaller Fs (p = 0.048) than young tendons. Seven older adult tendons exhibited tendon thickening and increased signal intensity indicative of sub-clinical tendinopathy. This subset of tendons exhibited a 7% smaller bound water fraction (p = 0.02) and significantly longer [Formula: see text] (p < 0.001) than the normal tendons from young and older adults.
CONCLUSION: Older adult tendons exhibited unique UTE signatures that are consistent with disruption of the collagen fiber network and changes in macromolecular content. UTE imaging metrics were sensitive to early indicators of tissue degeneration identified on anatomical images and hence could provide a quantitative biomarker by which to track changes in tissue health resulting from injury, disease, and treatment.