双机器人X射线系统超高分辨率锥形束CT原型用于尸体腕关节成像的评估
研究对16例尸体供体腕关节分别行锥形束CT(CBCT)原型与384层多探测器CT(MDCT)扫描,采用辐射当量低剂量(CTDIvol 3.3 mGy)和全剂量(CTDIvol 13.8 mGy)方案,评估图像质量并量化松质骨信号强度分数。CBCT主观图像质量优于同等剂量MDCT,全剂量CBCT中间信号强度像素分数最低,低剂量CBCT与全剂量MDCT无显著差异,CBCT较高分辨率MDCT可实现高达75%的剂量降低。
要点
- 全剂量CBCT主观图像质量中位Likert评分为7/7(阅片者1/阅片者2),优于全剂量MDCT的6/6。
- 低剂量CBCT评分为5/6,优于低剂量MDCT的3/3(p<0.001)。
- 全剂量CBCT中间信号强度像素分数为0.57(四分位距0.13),低于全剂量MDCT的0.68(0.21)。
关键影像表现/征象
- 全剂量CBCT主观图像质量优于辐射当量全剂量MDCT(p≤0.03)。
- 低剂量CBCT主观图像质量优于辐射当量低剂量MDCT(p<0.001)。
- 全剂量CBCT中间信号强度像素分数为0.57(四分位距0.13),全剂量MDCT为0.68(0.21)。
- 低剂量CBCT中间信号强度像素分数为0.72(0.19),低剂量MDCT为0.80(0.15)。
- 低剂量CBCT与全剂量MDCT之间中间信号强度像素分数无显著差异。
诊断标准与关键数值
- 主观图像质量采用七点Likert量表评估,中位值全剂量CBCT为7/7(阅片者1/阅片者2),全剂量MDCT为6/6,低剂量CBCT为5/6,低剂量MDCT为3/3。
- 组内相关系数为0.936(95%CI 0.897-0.961;p<0.001),提示信度极好。
- 以中间信号强度像素分数作为骨微结构显示指标,像素越少提示显示越优。
临床应用与阅片要点
- CBCT原型在腕关节成像中可提供优于高分辨率MDCT的骨小梁与骨髓图像质量,并较高分辨率MDCT剂量降低高达75%。
- 低剂量CBCT的客观骨微结构显示与全剂量MDCT无显著差异。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: Cone-beam CT (CBCT) applications possess potential for dose reduction in musculoskeletal imaging. This study evaluates the ultra-high-resolution CBCT prototype of a twin robotic X-ray system in wrist examinations compared to high-resolution multidetector CT (MDCT).
MATERIALS AND METHODS: Sixteen wrists of body donors were examined with the CBCT scan mode and a 384 slice MDCT system. Radiation-equivalent low-dose (CTDIvol(16cm) = 3.3 mGy) and full-dose protocols (CTDIvol(16cm) = 13.8 mGy) were used for both systems. Two observers assessed image quality on a seven-point Likert scale. In addition, software-assisted quantification of signal intensity fractions in cancellous bone was performed. Fewer pixels with intermediate signal intensity were considered to indicate superior depiction of bone microarchitecture.
RESULTS: Subjective image quality in CBCT was superior to dose equivalent MDCT with p ≤ 0.03 for full-dose and p < 0.001 for low-dose scans, respectively. Median Likert values were 7/7 (reader 1 / reader 2) in full-dose CBCT, 6/6 in full-dose MDCT, 5/6 in low-dose CBCT and 3/3 in low-dose MDCT. Intraclass correlation coefficient was 0.936 (95% confidence interval, 0.897-0.961; p < 0.001), indicating excellent reliability. Objective analysis displayed smaller fractions of "indecisive" pixels with intermediate signal intensity for full-dose CBCT (0.57 [interquartile range 0.13]) compared to full-dose MDCT (0.68 [0.21]), low-dose CBCT (0.72 [0.19]), and low-dose MDCT (0.80 [0.15]) studies. No significant difference was observed between low-dose CBCT and full-dose MDCT.
CONCLUSION: The new CBCT prototype provides superior image quality for trabecula and bone marrow in cadaveric wrist studies and enables dose reduction up to 75% compared to high-resolution MDCT.
引用
- Grunz JP, Weng AM, Gietzen CH, et al. Evaluation of Ultra-High-Resolution Cone-Beam CT Prototype of Twin Robotic Radiography System for Cadaveric Wrist Imaging. Acad Radiol. 2020. PMID: 32654956. doi:10.1016/j.acra.2020.06.018
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。