文献速递乳腺US

剪切波弹性成像与超微血管成像定量参数联合评估乳腺肿块

Combination of Quantitative Parameters of Shear Wave Elastography and Superb Microvascular Imaging to Evaluate Breast Masses.
PubMed 文献 · Korean J Radiol · 2020-01-01

回顾性分析192例患者200个病理证实的乳腺病灶,评估剪切波弹性成像(SWE)最大弹性值、比值及超微血管成像(SMI)血管指数联合B型超声鉴别乳腺肿块良恶性的价值。联合全部定量参数时AUC最高(0.849),准确率和特异度显著提高,敏感度无统计学下降。

要点

  • SWE最大弹性值界值86.45 kPa的AUC高于弹性比值3.57和SMI血管指数3.35%。
  • B型超声联合最大弹性值或SMI血管指数较单用B型超声AUC显著提高。
  • 联合最大弹性值、弹性比值和SMI血管指数时AUC达0.849,准确率由68.0%升至84.0%。

核心结论

  • 联合SWE和SMI全部定量参数与B型超声可提高鉴别乳腺病灶良恶性的诊断性能。

研究设计与样本

  • 回顾性分析192例患者共200个病理证实的乳腺病灶,采用B型超声、SWE和SMI检查,以病理为参考标准。

关键影像表现/征象

  • B型超声依据BI-RADS评估乳腺肿块。
  • SWE定量参数包括最大弹性值(Emax)和弹性比值(Eratio)。
  • SMI定量参数为血管指数(SMIVI)。

诊断标准与关键数值

  • Emax界值86.45 kPa的AUC最高,高于Eratio界值3.57和SMIVI界值3.35%。
  • 方案1中B型超声联合Emax或SMIVI的AUC分别为0.829和0.778,均高于单用B型超声的0.719(p < 0.001和p = 0.047)。
  • B型超声联合Emax、Eratio和SMIVI时AUC为0.849,为最佳诊断性能。
  • 联合全部定量参数时准确率由68.0%升至84.0%(p < 0.001),特异度由46.1%升至79.1%(p < 0.001)。
  • 联合全部定量参数时敏感度由97.6%降至90.6%,无统计学下降(p = 0.199)。

临床应用与阅片要点

  • 在B型超声基础上联合SWE最大弹性值和SMI血管指数可提高鉴别乳腺肿块良恶性的AUC。
  • 联合SWE最大弹性值、弹性比值和SMI血管指数可提高准确率和特异度,敏感度无统计学下降。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: This study aimed to evaluate the diagnostic value of combining the quantitative parameters of shear wave elastography (SWE) and superb microvascular imaging (SMI) to breast ultrasound (US) to differentiate between benign and malignant breast masses.

MATERIALS AND METHODS: A total of 200 pathologically confirmed breast lesions in 192 patients were retrospectively reviewed using breast US with B-mode imaging, SWE, and SMI. Breast masses were assessed based on the breast imaging reporting and data system (BI-RADS) and quantitative parameters using the maximum elasticity (Emax) and ratio (Eratio) in SWE and the vascular index in SMI (SMIVI). The area under the receiver operating characteristic curve (AUC) value, sensitivity, specificity, accuracy, negative predictive value, and positive predictive value of B-mode alone versus the combination of B-mode US with SWE or SMI of both parameters in differentiating between benign and malignant breast masses was compared, respectively. Hypothetical performances of selective downgrading of BI-RADS category 4a (set 1) and both upgrading of category 3 and downgrading of category 4a (set 2) were calculated.

RESULTS: Emax with a cutoff value of 86.45 kPa had the highest AUC value compared to Eratio of 3.57 or SMIVI of 3.35%. In set 1, the combination of B-mode with Emax or SMIVI had a significantly higher AUC value (0.829 and 0.778, respectively) than B-mode alone (0.719) (p < 0.001 and p = 0.047, respectively). B-mode US with the addition of Emax, Eratio, and SMIVI had the best diagnostic performance of AUC value (0.849). The accuracy and specificity increased significantly from 68.0% to 84.0% (p < 0.001) and from 46.1% to 79.1% (p < 0.001), respectively, and the sensitivity decreased from 97.6% to 90.6% without statistical loss (p = 0.199).

CONCLUSION: Combining all quantitative values of SWE and SMI with B-mode US improved the diagnostic performance in differentiating between benign and malignant breast lesions.

来源:Korean J Radiol,查看原文。版权归原作者所有。

引用

  1. Lee EJ, Chang YW. Combination of Quantitative Parameters of Shear Wave Elastography and Superb Microvascular Imaging to Evaluate Breast Masses. Korean J Radiol. 2020. PMID: 32691540. doi:10.3348/kjr.2019.0765

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。