健康志愿者肝脏剪切波弹性成像的观察者间与观察者内信度及空腹与非空腹状态的影响

Interobserver and Intraobserver Reliability of Hepatic Shear Wave Elastography and the Influence of Fasted Versus Nonfasted States in Healthy Volunteers.
PubMed 文献 · J Ultrasound Med · 2020-07-20

这项前瞻性研究纳入20名健康参与者,在空腹、饮水后及进食水后行肝脏剪切波弹性成像(SWE),评估空腹与进食对测值的影响及观察者间/内信度。饮水对SWE测值无显著影响;餐后1小时SWE测值有统计学显著变化(P=.04),3小时无显著变化(P=.08)。观察者间信度差至中等,观察者内信度极低至中等。

要点

  • 健康人饮水后肝脏SWE测值与空腹状态无显著差异。
  • 餐后1小时肝脏SWE测值有统计学显著变化(P=.04),3小时无显著变化(P=.08)。
  • 观察者间信度差至中等,观察者内信度极低至中等。

诊断标准与关键数值

  • 饮水后与空腹状态肝脏SWE测值无显著差异,统计学显著性设定为P<0.05。
  • 餐后状态在1小时对肝脏SWE测值有统计学显著影响(P=.04),3小时无显著影响(P=.08)。
  • 观察者间信度差至中等,观察者内信度极低至中等。
  • 观察者间SWE测值差值中位数为0.66至0.96 kPa。
  • 观察者内SWE测值差值中位数为0.43至0.55 kPa。

临床应用与阅片要点

  • 作者结论显示健康参与者饮水对肝脏SWE测值无影响。
  • 作者结论显示餐后1小时肝脏SWE测值有显著变化,3小时无显著变化。
  • 作者结论显示观察者间与观察者内一致性可变,最高仅为中等。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The primary objective of this study was to assess the effect of fasting versus ingestion of food and water on hepatic measurements by shear wave elastography (SWE) in healthy participants. The secondary objective was to assess inter- and intra-reader reliability of hepatic elastography in healthy participants.

METHODS: Twenty healthy participants were enrolled in this prospective study and underwent quantitative SWE under fasting conditions and after the ingestion of water and food and water. Two blinded sonographers each independently performed a total of 6 sessions of hepatic SWE in each participant. Sessions 1 to 3 were performed on day 0 and sessions 4 to 6 on day 7. Statistical tests used included the Wilcoxon signed ranks test, the intraclass correlation coefficient, and Bland-Altman plots.

RESULTS: There were no significant differences in hepatic SWE measurements after the ingestion of water versus the fasting state. Statistical significance was assessed as P < 0.05. The postprandial status had a statistically significant effect on hepatic SWE measurements at 1 hour (P = .04) but not at 3 hours (P = .08). By the intraclass correlation coefficient, there was poor-to-moderate inter-reader agreement and minimal-to-moderate intra-reader agreement. The median inter-reader difference in SWE measurements ranged from 0.66 to 0.96 kPa. The median intra-reader difference ranged from 0.43 to 0.55 kPa.

CONCLUSIONS: Our study shows that the ingestion of water has no effect on hepatic SWE measurements in healthy participants. The postprandial state had a significant effect on SWE measurements at 1 hour after ingestion but not at 3 hours. The inter-reader and intra-reader agreements were variable and moderate at best.