孕妇CT检查中胎儿辐射剂量的评估——一种网络工具的开发与验证
研究通过750次蒙特卡罗模拟构建计算算法,并在29例孕妇(孕8-35周)CT数据上验证,用于评估CT检查中胎儿吸收剂量。算法平均误差11%,优于商用软件,并开发了免费网络界面(www.fetaldose.org),需输入孕周、容积CTDI、管电压和扫描区域。
要点
- 腹部CT孕妇胎儿剂量中位数2.7 mGy,与母体周长中度相关(r=0.69)。
- 算法估算胎儿剂量平均误差11%,比商用工具更准确。
- 网络工具需输入孕周、容积CTDI、管电压和扫描区域,可独立于厂商计算。
核心结论
- 开发并验证了一种用于评估孕妇CT检查中胎儿剂量的工具,以免费且用户友好的网络界面提供。
研究设计与样本
- 研究分两部分:第一部分在代表不同孕期阶段的体模上进行750次蒙特卡罗模拟,并用2个厂商CT扫描仪的CTDI测量验证MC代码;第二部分经机构审查委员会批准,在29例孕妇(孕8-35周)的CT数据上进行患者特异性MC模拟验证算法,并与商用软件比较。
诊断标准与关键数值
- MC模拟得到的加权CTDI与2台CT系统测量值高度一致,平均误差4%。
- 腹部CT孕妇胎儿剂量中位数为2.7 mGy,与母体周长中度相关(r=0.69)。
- 算法估算胎儿剂量平均误差11%,比商用工具更准确。
- 网络界面(www.fetaldose.org)计算胎儿剂量需输入孕周、容积CTDI、管电压和扫描区域。
临床应用与阅片要点
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Estimations of radiation dose absorbed by the fetus from computed tomography (CT) in pregnant patients is mandatory, but currently available methods are not feasible in clinical routine. The aims of this study were to develop and validate a tool for assessment of fetal dose from CT of pregnant patients and to develop a user-friendly web interface for fast fetal dose calculations.
METHODS: In the first study part, 750 Monte Carlo (MC) simulations were performed on phantoms representing pregnant patients at various gestational stages. The MC code simulating vendor-independent dose distributions was validated against CT dose index (CTDI) measurements performed on CT scanners of 2 vendors. The volume CTDI-normalized fetal dose values from MC simulations were used for developing the computational algorithm enabling fetal dose assessments from CT of various body regions at different exposure settings. In the institutional review board-approved second part, the algorithm was validated against patient-specific MC simulations performed on CT data of 29 pregnant patients (gestational ages 8-35 weeks) who underwent CT. Furthermore, the tool was compared with a commercially available software. A user-friendly web-based interface for fetal dose calculations was created.
RESULTS: Weighted CTDI values obtained from MC simulations were in excellent agreement with measurements performed on the 2 CT systems (average error, 4%). The median fetal dose from abdominal CT in pregnant patients was 2.7 mGy, showing moderate correlation with maternal perimeter (r = 0.69). The algorithm provided accurate estimates of fetal doses (average error, 11%), being more accurate than the commercially available tool. The web-based interface (www.fetaldose.org) enabling vendor-independent calculations of fetal doses from CT requires the input of gestational age, volume CTDI, tube voltage, and scan region.
CONCLUSIONS: A tool for fetal dose assessments from CT of pregnant patients was developed and validated being freely available on a user-friendly web interface.