基于超声剪切波弹性成像的肝纤维化分类图像处理流程
本研究开发了一种基于超声剪切波弹性成像(SWE)的自动化方法,用于分类肝纤维化分期≥F2,并与手动参考方法比较。自动化方法在527例受试者的5526幅SWE图像上,AUROC为0.93(95%CI 0.90-0.94),参考方法为0.69(95%CI 0.65-0.72);在95%敏感度下特异度为71% vs 5%,每决策所需图像数为4幅 vs 8幅以上。
要点
- 自动化方法分类≥F2的AUROC为0.93,显著高于参考方法的0.69。
- 在95%敏感度下,自动化方法特异度为71%,参考方法仅5%。
- 自动化方法每决策仅需4幅SWE图像,参考方法需8幅以上。
临床应用与阅片要点
- 自动化流程整合图像质量评估、ROI自动选取和基于机器学习的多图像SWE分类,用于≥F2纤维化分类。
- 与手动参考方法相比,自动化方法在维持95%敏感度的同时将特异度从5%提升至71%,并将所需图像数从8幅以上降至4幅。
- 原文结论认为该自动化方法有潜力减少临床工作流程,但未给出具体阅片操作建议。
局限性
- 数据来自单一站点单一系统,原文明确说明为单站点单系统。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
The purpose of this study was to develop an automated method for classifying liver fibrosis stage ≥F2 based on ultrasound shear wave elastography (SWE) and to assess the system's performance in comparison with a reference manual approach. The reference approach consists of manually selecting a region of interest from each of eight or more SWE images, computing the mean tissue stiffness within each of the regions of interest and computing a resulting stiffness value as the median of the means. The 527-subject database consisted of 5526 SWE images and pathologist-scored biopsies, with data collected from a single system at a single site. The automated method integrates three modules that assess SWE image quality, select a region of interest from each SWE measurement and perform machine learning-based, multi-image SWE classification for fibrosis stage ≥F2. Several classification methods were developed and tested using fivefold cross-validation with training, validation and test sets partitioned by subject. Performance metrics were area under receiver operating characteristic curve (AUROC), specificity at 95% sensitivity and number of SWE images required. The final automated method yielded an AUROC of 0.93 (95% confidence interval: 0.90-0.94) versus 0.69 (95% confidence interval: 0.65-0.72) for the reference method, 71% specificity with 95% sensitivity versus 5% and four images per decision versus eight or more. In conclusion, the automated method reported in this study significantly improved the accuracy for ≥F2 classification of SWE measurements as well as reduced the number of measurements needed, which has the potential to reduce clinical workflow.
引用
- Brattain LJ, Ozturk A, Telfer BA, et al. Image Processing Pipeline for Liver Fibrosis Classification Using Ultrasound Shear Wave Elastography. Ultrasound Med Biol. 2020. PMID: 32622685. doi:10.1016/j.ultrasmedbio.2020.05.016
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。