手持超声设备自动定量分析二尖瓣环收缩期位移的可行性与可靠性:一项初步研究
本研究在手持超声设备中实现自动测量二尖瓣环收缩期位移(MAPSE)的算法,对20例患者与高端超声心动图的M型和彩色组织多普勒测量进行对比。自动法较M型低估MAPSE约8%–10%(差值1.2±1.4 mm,P<.005),与彩色组织多普勒差异无统计学意义(0.8±1.8 mm,P=.073),组内和组间ICC均≥0.86。作者结论为该自动定量左室纵向功能的方法可行、可靠,可整合入手持超声设备。
要点
- 手持超声设备自动测量的MAPSE较M型低估约8%–10%,与彩色组织多普勒差异无统计学意义。
- 自动法与M型、彩色组织多普勒的组内相关系数分别为0.85和0.81,一致性良好。
- 各方法组内与组间ICC均≥0.86,提示重复性优秀。
核心结论
- 手持超声设备自动定量左室纵向功能的方法具有高可行性和可靠性。
- 与M型成像相比,自动法低估MAPSE约8%–10%,但与彩色组织多普勒成像的差异无统计学意义。
- 作者结论为该自动定量评估左室功能的方法可整合入手持超声设备。
研究设计与样本
- 初步研究,共纳入20例大学医院患者,由心脏科医生或超声技师操作。
- 同时使用高端超声诊断仪行标准超声心动图检查,心尖四腔心切面分别用超声心动图和手持超声设备各记录4次。
- MAPSE在超声心动图上用M型和彩色组织多普勒测量,在手持超声设备上由算法自动测量。
诊断标准与关键数值
- 自动法MAPSE均值±标准差为9.6±2.2 mm,M型为10.9±2.6 mm,差值1.2±1.4 mm(P<.005)。
- 自动法与彩色组织多普勒的差值为0.8±1.8 mm(P=.073),差异无统计学意义。
- 自动法与M型测量的组内相关系数为0.85,与彩色组织多普勒测量为0.81。
- 各方法的组内与组间组内相关系数均≥0.86。
临床应用与阅片要点
- 原文未据此给出临床推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Handheld ultrasound devices (HUDs) have previously been limited to grayscale imaging without options for left ventricle (LV) quantification. We aimed to study the feasibility and reliability of automatic measurements of mitral annular plane systolic excursion (MAPSE) by HUDs.
METHODS: An algorithm that automatically measured MAPSE from live grayscale recordings was implemented in a HUD. Twenty patients at a university hospital were examined by either a cardiologist or a sonographer. Standard echocardiography using a high-end scanner was performed. The apical 4-chamber view was recorded 4 times by both echocardiography and the HUD. MAPSE was measured by M-mode and color tissue Doppler (cTD) during echocardiography and automatically by the HUD.
RESULTS: The automatic method underestimated mean MAPSE ± SD versus M-mode (9.6 ± 2.2 versus 10.9 ± 2.6 mm; difference, 1.2 ± 1.4 mm, P < .005). The difference between the automatic and cTD measurements was not significant (0.8 ± 1.8 mm; P = .073). The intraclass correlation coefficients (ICCs) between automatic and M-mode measurements was 0.85, and 0.81 for cTD measurements. There was good agreement between the methods, and the intra- and inter-rater ICCs were excellent for all methods (≥0.86).
CONCLUSIONS: In this novel study evaluating automatic quantification of LV longitudinal function by HUD, we showed the high feasibility and reliability of the method. Compared to M-mode imaging, the automatic method underestimated MAPSE by 8% to 10%, but the difference with cTD imaging was nonsignificant. We conclude that this study's method for automatic quantitative assessment of LV function can be integrated in HUDs.
引用
- Magelssen MI, Palmer CL, Hjorth-Hansen A, et al. Feasibility and Reliability of Automatic Quantitative Analyses of Mitral Annular Plane Systolic Excursion by Handheld Ultrasound Devices: A Pilot Study. J Ultrasound Med. 2020. PMID: 32710577. doi:10.1002/jum.15408
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。