四维血流涡量可视化预测定量血流测量不一致区域以优化血流测量
本回顾性研究纳入35例在3.0 T行四维血流MRI的患者,用基于λ2的涡量可视化与定量技术评估其能否识别定量血流不一致区域。升主动脉心输出量标准差在中等或重度涡量组为0.73 L/min ± 0.55,高于无或轻度涡量组的0.44 L/min ± 0.26(P < .01)。作者认为涡量可视化可能有助于指导血流定量的最佳位置。
要点
- 升主动脉心输出量标准差在中等或重度涡量组高于无或轻度涡量组(P < .01)。
- 全部患者升主动脉心输出量标准差高于肺动脉、降主动脉与上腔静脉(P < .001)。
- 涡量可视化可能有助于指导血流定量的最佳位置。
关键影像表现/征象
- 35例患者视觉分类为无涡量9例、轻度涡量8例、中度涡量11例、重度涡量7例。
- 全部患者升主动脉心输出量标准差为0.58 L/min ± 0.45,显著高于肺动脉的0.15 L/min ± 0.10以及降主动脉与上腔静脉的0.14 L/min ± 0.12(P < .001)。
- 升主动脉心输出量标准差在中等或重度涡量患者为0.73 L/min ± 0.55,显著高于无或轻度涡量患者的0.44 L/min ± 0.26(P < .01)。
诊断标准与关键数值
- 涡量视觉分级:无(n = 9)、轻度(n = 8)、中度(n = 11)、重度(n = 7)。
- 升主动脉瘤定义为最大直径 > 4.0 cm,18例存在。
- 10例存在二叶式主动脉瓣。
- 患者年龄58.6岁 ± 14.4(标准差),12例为女性。
临床应用与阅片要点
- 作者认为涡量可视化可能有助于指导血流定量的最佳位置。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To evaluate whether automated vorticity mapping four-dimensional (4D) flow MRI can identify regions of quantitative flow inconsistency.
MATERIALS AND METHODS: In this retrospective study, 35 consecutive patients who underwent MR angiography with 4D flow MRI at 3.0 T from December 2017 to October 2018 were analyzed using a λ 2-based technique for vorticity visualization and quantification. The patients were aged 58.6 years ± 14.4 (standard deviation), 12 were women, 18 had ascending aortic aneurysms (maximal diameter > 4.0 cm), and 10 had bicuspid aortic valves. Flow measurements were made in the ascending aorta (aAo), mid-descending aorta, main pulmonary artery, and superior vena cava. Statistical tests included t tests and F tests with a type I error threshold (α) of .05.
RESULTS: The 35 patients were visually classified as having no (n = 9), mild (n = 8), moderate (n = 11), or severe vorticity (n = 7). Across all patients, standard deviation of cardiac output in the aAo (0.58 L/min ± 0.45) was significantly (P < .001) higher than in the pulmonary arteries (0.15 L/min ± 0.10) and descending aorta and superior vena cava (0.14 L/min ± 0.12). The standard deviation of cardiac output observed in the aAo was significantly greater (P < .01) in patients with moderate or severe vorticity (0.73 L/min ± 0.55) than in those with none or mild vorticity (0.44 L/min ± 0.26).
CONCLUSION: Cardiac output and blood flow are essential MRI measurements in the evaluation of structural heart disease. Vorticity visualization may be used to help guide optimal location for flow quantification.© RSNA, 2020See also the commentary by Markl in this issue.