前视二维阵列对血流速度场的三维血管内表征
本研究使用118阵元前视环形阵列换能器与研究用超声系统,基于未聚焦波多普勒与横向振荡方法,在体模和猪股动脉中实现血管内三维血流速度矢量成像,并在狭窄体模中评估导管对血流动力学的影响,发现空心导管设计对狭窄处血流测量干扰较小。
要点
- 前视环形阵列结合未聚焦波多普勒与横向振荡可估计血管内三维血流速度矢量。
- 在狭窄体模中,空心导管设计对峰值流速测量干扰较小(误差6.5%)。
- 该技术未来可能用于已接受导管检查患者近壁三维血流动力学的评估。
关键影像表现/征象
- 在体模实验和猪股动脉在体实验中获取了实时三维血流速度矢量数据集。
- 血流速度分量在主要血流方向采用未聚焦波多普勒方法估计,在侧向和仰角方向采用横向振荡方法估计。
- 在狭窄体模中,峰值流速无导管时为54.97 ± 2.13 cm/s,有空心导管时为51.37 ± 1.08 cm/s,误差6.5%。
临床应用与阅片要点
- 作者指出,未来此类技术可能使已接受导管检查的患者能够评估近壁三维血流动力学。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Risk stratification in coronary artery disease is an ongoing challenge for which few tools are available for quantifying physiology within coronary arteries. Recently, anatomy-driven computational fluid dynamic modeling has enabled the mapping of local flow dynamics in coronary stenoses, with derived parameters such as WSS exhibiting a strong capability for predicting adverse clinical events on a patient-specific basis. As cardiac catheterization is common in patients with coronary artery disease, minimally invasive technologies capable of identifying pathologic flow in situ in real time could have a significant impact on clinical decision- making. As a step toward in vivo quantification of slow flow near the arterial wall, proof-of-concept for 3-D intravascular imaging of blood flow dynamics is provided using a 118-element forward-viewing ring array transducer and a research ultrasound system. Blood flow velocity components are estimated in the direction of primary flow using an unfocused wave Doppler approach, and in the lateral and elevation directions, using a transverse oscillation approach. This intravascular 3-D vector velocity system is illustrated by acquiring real-time 3-D data sets in phantom experiments and in vivo in the femoral artery of a pig. The effect of the catheter on blood flow dynamics is also experimentally assessed in flow phantoms with both straight and stenotic vessels. Results indicate that 3-D flow dynamics can be measured using a small form factor device and that a hollow catheter design may provide minimal disturbance to flow measurements in a stenosis (peak velocity: 54.97 ± 2.13 cm/s without catheter vs. 51.37 ± 1.08 cm/s with hollow catheter, 6.5% error). In the future, such technologies could enable estimation of 3-D flow dynamics near the wall in patients already undergoing catheterization.
引用
- Lindsey BD, Jing B, Kim S, et al. 3-D Intravascular Characterization of Blood Flow Velocity Fields with a Forward-Viewing 2-D Array. Ultrasound Med Biol. 2020. PMID: 32616428. doi:10.1016/j.ultrasmedbio.2020.05.022
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。