新型装置实现胎儿内镜气管封堵的无创逆转:离体与在体研究

New device permitting non-invasive reversal of fetal endoscopic tracheal occlusion: ex-vivo and in-vivo study.
PubMed 文献 · Ultrasound Obstet Gynecol · 2020-09-11

研究评估Smart-TO球囊在模拟宫内环境及胎羊模型中经1.5-T磁共振暴露实现气管封堵逆转的可行性、闭塞效果与局部反应。离体模拟中多数体位单次MR暴露即放气,仅坐轮椅体位出现1例三次暴露后仍未放气;胎羊在体所有Smart-TO球囊均成功放气,肺体重量比显著高于未封堵对照且与Goldbal2组相当,气管组织学改变轻微。

要点

  • Smart-TO球囊可在1.5-T MR磁场暴露下自发放气,离体模拟中仅坐轮椅体位出现放气失败。
  • 胎羊在体实验中Smart-TO球囊全部成功放气,肺体重量比显著高于未封堵对照。
  • Smart-TO与临床标准Goldbal2在肺形态计量、气管周径及气管组织学上无差异。

核心结论

  • 在模拟宫内环境中,胎儿处于不同体位时经1.5-T MR系统磁场暴露可有效使Smart-TO球囊放气,仅母亲坐轮椅体位时出现1例失败。
  • 在健康胎羊中,Smart-TO球囊的闭塞效果与临床标准Goldbal2系统相当,局部副作用有限。

研究设计与样本

  • 离体研究:在高保真模拟器中经视频镜置入Smart-TO球囊,将胎儿模型置于1 L充水球囊内模拟羊膜腔,由操作者持于腹前模拟不同胎儿与母亲体位,经1.5-T MR机磁场暴露后行气管镜评估放气,失败者最多重复MR暴露三次。
  • 在体研究:胎羊于妊娠第95天(足月145天)行气管封堵,Goldbal2组n=5、Smart-TO组n=5,另设未封堵对照6只;妊娠第116天经气管镜评估球囊,Goldbal2以穿刺放气、Smart-TO以MR暴露放气,安乐死后评估肺体重量比、肺形态计量及气管周径,并以分级组织学评分系统评估气管局部改变。

关键影像表现/征象

  • 离体模拟中母亲站立位、模型高度95 cm(n=32)、55 cm(n=8)或125 cm(n=8)以及母亲「平卧于担架」位(n=8)时,单次MR暴露后Smart-TO球囊放气率为100%。
  • 离体模拟中母亲「坐于轮椅」位时8个球囊中3个(37.5%)首次暴露未放气,其中2个经第二次MR暴露后放气,1个经第三次暴露后仍保持充气。
  • 胎羊在体实验中所有Smart-TO球囊均成功放气;Smart-TO组肺体重量比显著高于未封堵对照组,且与Goldbal2组相当。
  • 两种球囊在肺形态计量与气管周径上无差异,气管组织学均显示轻微改变。

局限性

  • 原文结论指出该装置的安全性与有效性尚未得到证实。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: One of the drawbacks of fetal endoscopic tracheal occlusion (FETO) for congenital diaphragmatic hernia is the need for a second invasive intervention to re-establish airway patency. The 'Smart-TO' device is a new balloon for FETO that deflates spontaneously when placed in a strong magnetic field, therefore overcoming the need for a second procedure. The safety and efficacy of this device have not yet been demonstrated. The aim of this study was to investigate the reversibility, local side effects and occlusiveness of the Smart-TO balloon, both in a simulated in-utero environment and in the fetal lamb model.

METHODS: First, the reversibility of tracheal occlusion by the Smart-TO balloon was tested in a high-fidelity simulator. Following videoscopic tracheoscopic balloon insertion, the fetal mannequin was placed within a 1-L water-filled balloon to mimic the amniotic cavity. This was held by an operator in front of their abdomen, and different fetal and maternal positions were simulated to mimic the most common clinical scenarios. Following exposure to the magnetic field generated by a 1.5-T magnetic resonance (MR) machine, deflation of the Smart-TO balloon was assessed by tracheoscopy. In cases of failed deflation, the mannequin was reinserted into a water-filled balloon for additional MR exposure, up to a maximum of three times. Secondly, reversibility, occlusiveness and local effects of the Smart-TO balloon were tested in vivo in fetal lambs. Tracheal occlusion was performed in fetal lambs on gestational day 95 (term, 145 days), either using the balloon currently used in clinical practice (Goldbal2) (n = 5) or the Smart-TO balloon (n = 5). On gestational day 116, the presence of the balloon was assessed by tracheoscopy. Deflation was performed by puncture (Goldbal2) or MR exposure (Smart-TO). Six unoccluded fetal lambs served as controls. Following euthanasia, the lung-to-body-weight ratio (LBWR), lung morphometry and tracheal circumference were assessed. Local tracheal changes were measured using a hierarchical histologic scoring system.

RESULTS: Ex vivo, Smart-TO balloon deflation occurred after a single MR exposure in 100% of cases in a maternal standing position with the mannequin at a height of 95 cm (n = 32), 55 cm (n = 8) or 125 cm (n = 8), as well as when the maternal position was 'lying on a stretcher' (n = 8). Three out of eight (37.5%) balloons failed to deflate at first exposure when the maternal position was 'sitting in a wheelchair'. Of these, two balloons deflated after a second MR exposure, but one balloon remained inflated after a third exposure. In vivo, all Smart-TO balloons deflated successfully. The LBWR in fetal lambs with tracheal occlusion by a Smart-TO balloon was significantly higher than that in unoccluded controls, and was comparable with that in the Goldbal2 group. There were no differences in lung morphometry and tracheal circumference between the two balloon types. Tracheal histology showed minimal changes for both balloons.

CONCLUSIONS: In a simulated in-utero environment, the Smart-TO balloon was effectively deflated by exposure of the fetus in different positions to the magnetic field of a 1.5-T MR system. There was only one failure, which occurred when the mother was sitting in a wheelchair. In healthy fetal lambs, the Smart-TO balloon is as occlusive as the clinical standard Goldbal2 system and has only limited local side effects. © 2020 International Society of Ultrasound in Obstetrics and Gynecology.

来源:Ultrasound Obstet Gynecol,查看原文。版权归原作者所有。