文献速递神经技术与安全CT

成人双源CT非增强头颅虚拟单能重建图像质量评估

Image Quality of Virtual Monochromatic Reconstructions of Noncontrast CT on a Dual-Source CT Scanner in Adult Patients.
PubMed 文献 · Acad Radiol · 2020-06-30

回顾性重建25例成人双源非增强头颅CT的40–140 keV虚拟单能图像,采用定量核与头颅核评估CT值、噪声、SNR、CNR及后颅窝伪影指数。CT值随能量降低而升高,SD在85 keV最低,SNR在头颅核80 keV与定量核85 keV最高,CNR在40 keV最高,PFAI在头颅核90 keV与定量核100 keV最低。最佳虚拟单能图像质量显著优于常规CT,但取决于重建核与检查指征。

要点

  • 虚拟单能图像CT值随能量降低而升高,噪声SD在85 keV最低。
  • SNR在头颅核80 keV与定量核85 keV最高,CNR在40 keV最高。
  • 后颅窝伪影指数PFAI在头颅核90 keV与定量核100 keV最低。

核心结论

  • 最佳虚拟单能图像能量可改善脑实质图像质量,且显著优于常规CT,但取决于重建核与进行非增强CT的指征。

研究设计与样本

  • 回顾性重建研究,纳入25例连续成人患者,均接受双源双能量非增强头颅CT。
  • 以5 keV为间隔在40至140 keV范围重建虚拟单能图像,采用定量核与头颅核。
  • 评估指标包括CT值、噪声(SD)、信噪比(SNR)、对比噪声比(CNR)及后颅窝伪影指数(PFAI),并与常规CT比较。

关键影像表现/征象

  • CT值随虚拟单能能量水平降低而升高,SD在85 keV最低。
  • SNR在头颅核80 keV和定量核85 keV分别达到最大。
  • CNR在40 keV最大;PFAI在头颅核90 keV和定量核100 keV分别最低。
  • 最佳虚拟单能图像质量显著优于常规CT。

诊断标准与关键数值

  • SNR最大值对应能量:头颅核80 keV,定量核85 keV。
  • CNR最大值对应能量:40 keV。
  • PFAI最低值对应能量:头颅核90 keV,定量核100 keV。
  • SD最低值对应能量:85 keV。

临床应用与阅片要点

  • 虚拟单能图像质量优化需结合重建核与临床指征选择能量水平,最佳能量因重建核而异。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: To evaluate the image quality of virtual monochromatic images (VMI) reconstructed from dual-energy dual-source noncontrast head CT with different reconstruction kernels.

MATERIALS AND METHODS: Twenty-five consecutive adult patients underwent noncontrast dual-energy CT. VMI were retrospectively reconstructed at 5-keV increments from 40 to 140 keV using quantitative and head kernels. CT-number, noise levels (SD), signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) in the gray and white matter and artifacts using the posterior fossa artifact index (PFAI) were evaluated.

RESULTS: CT-number increased with decreasing VMI energy levels, and SD was lowest at 85 keV. SNR was maximized at 80 keV and 85 keV for the head and quantitative kernels, respectively. CNR was maximum at 40 keV; PFAI was lowest at 90 (head kernel) and 100 (quantitative kernel) keV. Optimal VMI image quality was significantly better than conventional CT.

CONCLUSION: Optimal image quality of VMI energies can improve brain parenchymal image quality compared to conventional CT but are reconstruction kernel dependent and depend on indication for performing noncontrast CT.

来源:Acad Radiol,查看原文。版权归原作者所有。