在新生儿重症监护病房使用新型移动数字X线摄影系统开发质控低剂量摄影方案
前瞻性随机研究纳入40例NICU新生儿,比较常规技术与新型移动DR系统三种低剂量方案(A等效100%、B等效80%、C等效64%剂量)的胸腹部摄影图像质量;结果显示方案B和C对常规方案在整体图像质量上非劣效,方案B对几乎全部解剖标志的显示及整体图像质量优于常规方案。
要点
- 低剂量方案B(80%等效剂量)和C(64%等效剂量)整体图像质量对常规方案非劣效。
- 方案B对几乎全部解剖标志的显示及整体图像质量优于常规方案。
- 三位阅片者间信度0.91(p < 0.001)。
关键影像表现/征象
- 方案B对几乎全部解剖标志的显示优于常规方案。
- 方案B的整体图像质量优于常规方案。
诊断标准与关键数值
- 三位阅片者间信度为0.91(p < 0.001)。
- 方案B和方案C对常规方案在整体图像质量上均达到统计学非劣效。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE. The aim of this study was to develop a low-dose radiography protocol for the neonatal ICU (NICU) using a new mobile digital radiography system with advanced denoising image processing and to evaluate the noninferiority of that protocol. SUBJECTS AND METHODS. In this prospective randomized study, 40 neonates in the NICU underwent radiography of the thorax and abdomen with two different mobile radiography units: conventional technique on one unit (50 kV, 1.6 mAs, and no additional filtration) and a new technique on another unit (54 kV, 0.1-mm Cu filtration). Three low-dose protocols for the second unit were developed in a phantom study: protocol A (100% equivalent dose with conventional protocol), protocol B (80% equivalent dose), and protocol C (64% equivalent dose). The noninferiority of each low-dose protocol was assessed by three independent readers using image quality criteria. RESULTS. Forty patients each underwent three pairs of radiography examinations (protocol A and the conventional protocol, protocol B and the conventional protocol, and protocol C and the conventional protocol), except one pair that did not include one image of the conventional protocol. The interrater reliability among the three readers was 0.91 (p < 0.001). Both of the low-dose protocols (B and C) were statistically noninferior to the conventional protocol with respect to overall image quality. Protocol B better depicted almost all anatomic landmarks and had better overall image quality than the conventional protocol. CONCLUSION. Using appropriate technique and acquisition factors, radiation dose can be lowered on a digital radiography system without significant effect on the image quality by adding filtrations and a new denoising technique.
引用
- Choi G, Cheon JE, Lee S, et al. Development of Quality-Controlled Low-Dose Protocols for Radiography in the Neonatal ICU Using a New Mobile Digital Radiography System. AJR Am J Roentgenol. 2020. PMID: 32406770. doi:10.2214/AJR.19.22269
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。