1.5 T 磁共振金属伪影校正技术诊断全髋关节置换术后假体周围关节感染

Diagnosis of Periprosthetic Hip Joint Infection Using MRI with Metal Artifact Reduction at 1.5 T.
PubMed 文献 · Radiology · 2020-05-12

回顾性评估2015年1月至2018年4月间40例全髋关节置换术后假体周围关节感染患者与100例无感染对照的1.5 T磁共振金属伪影校正图像。骨膜反应、关节囊水肿和肌肉内水肿在两组间差异有统计学意义(P < .001),诊断准确率分别为86%、91%和89%;骨溶解、骨髓水肿、积液、外展肌腱病变及滑囊炎在两组间无差异(P > .05)。

要点

  • 骨膜反应、关节囊水肿、肌肉内水肿对假体周围关节感染有较高诊断准确率。
  • 肌肉内水肿敏感度最高(95%),关节囊水肿特异度最高(95%)。
  • 骨溶解、骨髓水肿、积液、外展肌腱病变及滑囊炎不能区分感染与非感染。

关键影像表现/征象

  • 骨膜反应在感染组40例中31例、对照组100例中10例出现。
  • 关节囊水肿在感染组40例中33例、对照组100例中5例出现。
  • 肌肉内水肿在感染组40例中38例、对照组100例中14例出现。
  • 骨溶解在感染组40例中23例、对照组100例中60例,骨髓水肿在感染组40例中39例、对照组100例中87例,积液在感染组40例中20例、对照组100例中26例,外展肌腱病变在感染组40例中22例、对照组100例中62例,滑囊炎在感染组40例中14例、对照组100例中34例,两组间差异均无统计学意义(P > .05)。

诊断标准与关键数值

  • 骨膜反应诊断假体周围关节感染的敏感度78%、特异度90%、准确率86%。
  • 关节囊水肿诊断假体周围关节感染的敏感度83%、特异度95%、准确率91%。
  • 肌肉内水肿诊断假体周围关节感染的敏感度95%、特异度86%、准确率89%。
  • 观察者间信度几乎完美,κ值为0.88至0.92。

临床应用与阅片要点

  • 阅读全髋关节置换术后1.5 T磁共振金属伪影校正图像时,可关注骨膜反应、关节囊水肿和肌肉内水肿以辅助诊断假体周围关节感染。
  • 骨溶解、骨髓水肿、积液、外展肌腱病变及滑囊炎在感染与非感染患者中均常见,单独出现时鉴别价值有限。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Background MRI with metal artifact reduction has gained importance in assessment of pain with total hip arthroplasty (THA). However, its role in diagnosis of periprosthetic joint infection (PJI) has not been well defined. Purpose To evaluate findings of PJI after THA and to determine the diagnostic performance of 1.5-T MRI with metal artifact reduction. Materials and Methods Dedicated compressed sensing-based slice encoding for metal artifact correction 1.5-T MRI examinations (from January 2015 to April 2018) in patients with THA PJI (infection group) and noninfected THA (control group) were retrospectively evaluated by two musculoskeletal radiologists. Fisher exact test was used to compare the groups. Sensitivity, specificity, and accuracy were evaluated for each finding. Interobserver reliability was assessed with κ statistics. Results Forty patients (mean age, 69 years ± 11 [standard deviation]; 26 men) in the infection group and 100 patients (mean age, 67 years ± 11; 42 men) in the control group were evaluated. Periosteal reaction, capsule edema, and intramuscular edema differed between the two groups (P < .001 for each finding). Periosteal reaction was found in 31 of 40 patients with infection and 10 of 100 participants in the control group (sensitivity, 78%; specificity, 90%; accuracy, 86%); capsule edema was found in 33 of 40 (infection group) and five of 100 (control group) (sensitivity, 83%; specificity, 95%; accuracy, 91%); and intramuscular edema was found in 38 of 40 (infection group) and 14 of 100 (control group) (sensitivity, 95%; specificity, 86%; accuracy, 89%). Interobserver agreement was almost perfect, with κ values between 0.88 and 0.92. No difference between the two groups was found regarding the presence of osteolysis (infection group, 23 of 40; control group, 60 of 100), bone marrow edema (39 of 40 vs 87 of 100), effusion (20 of 40 vs 26 of 100), abductor tendon lesion (22 of 40 vs 62 of 100), or bursitis (14 of 40 vs 34 of 100) (P > .05 for each finding). Conclusion The presence of periosteal reaction, capsule edema, and intramuscular edema after total hip arthroplasty at 1.5-T MRI with metal artifact reduction had a high accuracy in evaluation of periprosthetic joint infection. © RSNA, 2020 Online supplemental material is available for this article. See also the editorial by Zanetti in this issue.