膝关节超分辨率磁共振成像:采用二维涡轮自旋回波成像
本研究在3-T扫描仪上对体模和22名受试者(8名健康志愿者和14名患者)行二维涡轮自旋回波超分辨率重建与三维涡轮自旋回波SPACE采集,比较图像质量。结果显示超分辨率重建在体模中空间分辨率优于SPACE,在健康志愿者中信号噪声比和对比噪声比无显著差异,半月板可见性和诊断信心更高,但图像对比度和液体亮度较低,软骨可见性SPACE更高。
要点
- 超分辨率重建在体模中空间分辨率优于传统三维SPACE。
- 健康志愿者中信号噪声比和对比噪声比在超分辨率重建与SPACE间无显著差异。
- 半月板诊断信心超分辨率重建显著高于SPACE,软骨可见性SPACE显著更高。
关键影像表现/征象
- 体模中感知模糊度量和边缘响应函数显示超分辨率重建较传统三维SPACE空间分辨率明显改善。
- 健康志愿者临床相关结构中超分辨率重建与三维SPACE的信号噪声比和对比噪声比无显著差异。
- 超分辨率重建整体图像质量更好且边缘模糊更少,但感知图像对比度三维SPACE更好。
- 半月板可见性评分超分辨率重建显著更好,软骨可见性三维SPACE显著更高,韧带在两者上均高可见。
- 半月板诊断信心超分辨率重建显著高于三维SPACE,软骨和韧带诊断信心两者无显著差异。
诊断标准与关键数值
- 超分辨率重建采用7个各向异性二维涡轮自旋回波堆栈,体素0.5×0.5×2.0 mm,每堆栈扫描时间1分55秒,总扫描时间13分25秒,重建体素0.5×0.5×0.5 mm。
- 传统SPACE序列体素0.5×0.5×0.5 mm,扫描时间12分42秒。
- 半月板评估的读者间一致性三维SPACE为实质性,超分辨率重建为几乎完美;软骨评估的一致性三维SPACE为几乎完美,超分辨率重建为中等。
临床应用与阅片要点
- 超分辨率重建可提供各向同性高分辨率膝关节图像,阅片时需注意其图像对比度和液体亮度低于传统三维SPACE。
- 评估半月板时超分辨率重建的可见性和诊断信心更高,但评估软骨时三维SPACE可见性更高。
- 作者认为未来需进行对比度优化和采用先进加速技术缩短采集时间,以对超分辨率重建膝关节磁共振成像进行临床验证。
局限性
- 作者指出未来需进行对比度优化和采用先进加速技术缩短采集时间,以对超分辨率重建膝关节磁共振成像进行临床验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The purpose of this study was to assess the technical feasibility of 3-dimensional (3D) super-resolution reconstruction (SRR) of 2D turbo spin echo (TSE) knee magnetic resonance imaging (MRI) and to compare its image quality with conventional 3D TSE sampling perfection with application optimized contrast using different flip angle evolutions (SPACE) MRI.
MATERIALS AND METHODS: Super-resolution reconstruction 2D TSE MRI and 3D TSE SPACE images were acquired from a phantom and from the knee of 22 subjects (8 healthy volunteers and 14 patients) using a clinical 3-T scanner. For SRR, 7 anisotropic 2D TSE stacks (voxel size, 0.5 × 0.5 × 2.0 mm; scan time per stack, 1 minute 55 seconds; total scan time, 13 minutes 25 seconds) were acquired with the slice stack rotated around the phase-encoding axis. Super-resolution reconstruction was performed at an isotropic high-resolution grid with a voxel size of 0.5 × 0.5 × 0.5 mm. Direct isotropic 3D image acquisition was performed with the conventional SPACE sequence (voxel size, 0.5 × 0.5 × 0.5 mm; scan time, 12 minutes 42 seconds). For quantitative evaluation, perceptual blur metrics and edge response functions were obtained in the phantom image, and signal-to-noise and contrast-to-noise ratios were measured in the images from the healthy volunteers. Images were qualitatively evaluated by 2 independent radiologists in terms of overall image quality, edge blurring, anatomic visibility, and diagnostic confidence to assess normal and abnormal knee structures. Nonparametric statistical analysis was performed, and significance was defined for P values less than 0.05.
RESULTS: In the phantom, perceptual blur metrics and edge response functions demonstrated a clear improvement in spatial resolution for SRR compared with conventional 3D SPACE. In healthy subjects, signal-to-noise and contrast-to-noise ratios in clinically relevant structures were not significantly different between SRR and 3D SPACE. Super-resolution reconstruction provided better overall image quality and less edge blurring than conventional 3D SPACE, yet the perceived image contrast was better for 3D SPACE. Super-resolution reconstruction received significantly better visibility scores for the menisci, whereas the visibility of cartilage was significantly higher for 3D SPACE. Ligaments had high visibility on both SRR and 3D SPACE images. The diagnostic confidence for assessing menisci was significantly higher for SRR than for conventional 3D SPACE, whereas there were no significant differences between SRR and 3D SPACE for cartilage and ligaments. The interreader agreement for assessing menisci was substantial with 3D SPACE and almost perfect with SRR, and the agreement for assessing cartilage was almost perfect with 3D SPACE and moderate with SRR.
CONCLUSIONS: We demonstrate the technical feasibility of SRR for high-resolution isotropic knee MRI. Our SRR results show superior image quality in terms of edge blurring, but lower image contrast and fluid brightness when compared with conventional 3D SPACE acquisitions. Further contrast optimization and shortening of the acquisition time with state-of-the-art acceleration techniques are necessary for future clinical validation of SRR knee MRI.
引用
- Van Dyck P, Smekens C, Vanhevel F, et al. Super-Resolution Magnetic Resonance Imaging of the Knee Using 2-Dimensional Turbo Spin Echo Imaging. Invest Radiol. 2020. PMID: 32404629. doi:10.1097/RLI.0000000000000676