基于临床在体弹性成像数据的有限元建模用于乳腺肿瘤诊断
本研究利用有限元建模模拟超声弹性成像中乳腺组织的多次压缩,依据非线性生物力学特性对乳腺肿瘤分类。数值模拟先以虚拟模型验证,再基于在体数据构建个体化三维模型;良恶性组分类参数差异有统计学意义(P < .05),在体与有限元分类结果强相关(R = 0.875;P < .001)且无统计学差异(P = .909)。
要点
- 良性与恶性肿瘤的分类参数差异有统计学意义(P < .05)。
- 在体分类与有限元模型分类结果强相关(R = 0.875;P < .001)。
- 在体与有限元两种分类结果之间无统计学差异(P = .909)。
核心结论
- 模型结果与在体诊断之间一致性良好,所构建模型经过验证与确认。
- 所引入的三维建模方法可能增强弹性成像方法,用于在早期对乳腺肿瘤进行初步分类。
研究设计与样本
- 研究先以假定肿瘤几何尺寸但采用实际材料属性的三维虚拟模型进行数值模拟,以检验和验证有限元建模。
- 进一步基于在体实验数据构建三维模型以验证模型,模型按每个在体病例个体化设计,强调乳腺组织的几何形态、位置和生物力学属性。
- 在不同压缩水平下分析肿瘤与背景正常组织之间的组织应变,以探讨其非线性并分类肿瘤类型。
- 肿瘤分类参数通过施加的压缩力与应变差异之间的幂律关系推导。
关键影像表现/征象
- 在不同压缩水平下分析肿瘤与背景正常组织之间的组织应变,以探讨其非线性。
- 通过施加压缩力与应变差异之间的幂律关系推导肿瘤分类参数。
诊断标准与关键数值
- 良性与恶性肿瘤的分类参数在有限元验证和确认中差异均有统计学意义(P < .05)。
- 在体分类与有限元分类参数强相关(R = 0.875;P < .001)。
- 在体与有限元两种分类结果之间无统计学差异(P = .909)。
临床应用与阅片要点
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: This study exploited finite-element modeling (FEM) to simulate breast tissue multicompression during ultrasound elastography to classify breast tumors based on their nonlinear biomechanical properties.
METHODS: Numeric simulations were first calculated by using 3-dimensional (3D) virtual models with an assumed tumor's geometric dimensions but with actual material properties to test and validate the FEM. Further numeric simulations were used to construct 3D models based on in vivo experimental data to verify our models. The models were designed for each individual in vivo case, emphasizing the geometry, position, and biomechanical properties of the breast tissue. At different compression levels, tissue strains were analyzed between the tumors and the background normal tissues to explore their nonlinearity and classify the tumor type. Tumor classification parameters were deduced by using a power-law relationship between the applied compressive forces and strain differences.
RESULTS: Classification parameters were compared between benign and malignant tumors, for which they were found to be statistically significant in classifying the tumor types (P < .05) by both the validation and verification of FEM. We compared the classification parameters between the in vivo and FEM classifications, for which they were found to be strongly correlated (R = 0.875; P < .001), with no statistical differences between their outcomes (P = .909).
CONCLUSIONS: Good agreement between the model outcomes and the in vivo diagnostics was reported. The implemented models were validated and verified. The introduced 3D modeling method may augment elastographic methods to preliminary classify breast tumors at an early stage.
引用
- Sayed AM, Naser MA, Wahba AA, et al. Breast Tumor Diagnosis Using Finite-Element Modeling Based on Clinical in vivo Elastographic Data. J Ultrasound Med. 2020. PMID: 32472949. doi:10.1002/jum.15344
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。