带射频识别端口的乳腺组织扩张器:MRI相关问题评估

Breast Tissue Expander With Radiofrequency Identification Port: Assessment of MRI Issues.
PubMed 文献 · AJR Am J Roentgenol · 2020-05-14

本研究采用标准化方法评估带射频识别端口的乳腺组织扩张器(Motiva Flora)在3T磁场交互、1.5T/3T加热、3T伪影及功能变化方面的MRI安全性。结果显示磁场交互轻微、温度升高无生理后果、伪影较小,且不同MRI条件下端口定位和电子序列号读取不受影响,结论为该扩张器在1.5T或3T MRI条件下为「MR Conditional」。

要点

  • 带射频识别端口的乳腺组织扩张器在3T下偏转角仅2°且无扭矩,磁场交互轻微。
  • 1.5T/64MHz与3T/128MHz下最高温升分别为1.7°C和1.9°C,无生理后果。
  • 该扩张器在3T MRI上伪影相对较小,且不同MRI条件不影响端口定位与电子序列号读取。

核心结论

  • 带射频识别端口的乳腺组织扩张器在1.5T或3T MRI条件下为「MR Conditional」。
  • 与带磁性端口、在MR图像上产生明显信号丢失和畸变的组织扩张器相比,该植入物伪影相对较小,可能为接受MRI的患者带来潜在优势。

研究设计与样本

  • 研究评估带射频识别端口的乳腺组织扩张器(Motiva Flora Tissue Expander,Establishment Labs)的MRI相关问题。
  • 评估内容包括3T磁场交互(平移吸引力和扭矩)、1.5T/64MHz和3T/128MHz的MRI相关加热、3T伪影,以及不同MRI条件下(1.5T/64MHz和3T/128MHz)的功能变化。

关键影像表现/征象

  • 磁场交互轻微,偏转角为2°且无扭矩。
  • 3T MRI上该带射频识别端口组织扩张器表现为相对较小的伪影。

诊断标准与关键数值

  • 1.5T/64MHz下最高温度升高1.7°C,3T/128MHz下最高温度升高1.9°C。
  • 磁场交互评估中偏转角为2°且无扭矩。
  • 暴露于不同MRI条件后仍能定位射频识别端口并读取电子序列号。

临床应用与阅片要点

  • 该带射频识别端口的组织扩张器在1.5T或3T MRI条件下为「MR Conditional」。
  • 与带磁性端口的组织扩张器相比,该植入物在MR图像上伪影相对较小,可能为接受MRI的患者带来潜在优势。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE. Breast tissue expanders with magnetic ports are MRI unsafe, preventing patients from benefiting from the diagnostic capabilities of MRI. A tissue expander was recently developed with a radiofrequency identification (RFID) port used for needle location and expansion that may be acceptable for a patient undergoing MRI. The purpose of this investigation was to evaluate MRI issues using standardized techniques and well-accepted methods for this tissue expander with RFID port. MATERIALS AND METHODS. The breast tissue expander with RFID port (Motiva Flora Tissue Expander, Establishment Labs) was assessed for magnetic field interactions (translational attraction and torque, 3 T), MRI-related heating (1.5 T/64 MHz and 3 T/128 MHz), artifacts (3 T), and functional changes associated with different MRI conditions (1.5 T/64 MHz and 3 T/128 MHz). RESULTS. Magnetic field interactions were minor (deflection angle of 2° and no torque) and thus will not pose a risk. At 1.5 T/64 MHz and 3 T/128 MHz, the highest temperature elevations (1.7°C and 1.9°C, respectively) were physiologically inconsequential. The tissue expander with RFID port exhibited relatively small artifacts on MRI. Exposures of the tissue expander with RFID port to different MRI conditions did not impact the ability to localize the RFID port or to read the electronic serial number. CONCLUSION. The findings indicated that this tissue expander with RFID port is "MR Conditional" for a patient referred for MRI at 1.5 T or 3 T. Importantly, the relatively small artifact associated with this implant offers potential advantages for patients undergoing MRI compared with tissue expanders that have magnetic ports that create substantial signal losses and distortions on MR images.