双能CT虚拟单能成像能否改善脂肪肝患者低密度肝转移瘤的评估
本研究纳入28例脂肪肝伴低密度肝转移瘤患者,行平扫及门静脉期双能CT,重建标准线性融合与10-keV间隔虚拟单能图像。40-keV VMI的病灶-肝实质对比噪声比最高(4.3±2.6 vs 标准2.9±1.9,p<0.001),病灶边缘勾画主观评分更优(中位5分,p≤0.001),测量一致性极好。
要点
- 40-keV虚拟单能图像较标准重建显著提高低密度肝转移瘤的对比噪声比。
- 40-keV虚拟单能图像对病灶边缘勾画的主观评分优于标准重建。
- 两种重建方式下病灶测量的观察者间与观察者内一致性均极好。
核心结论
- 噪声优化的虚拟单能重建较标准重建显著改善脂肪肝患者低密度肝转移瘤的对比度和病灶边缘勾画。
研究设计与样本
- 研究类型为对28例脂肪肝伴低密度肝转移瘤患者的影像学评估,患者接受平扫及门静脉期双能CT。
- 重建标准线性融合图像及以10-keV为间隔的虚拟单能图像系列。
- 计算病灶-肝实质对比噪声比(CNR),并对最佳虚拟单能系列进行整体图像质量和病灶边缘勾画的主观评估。
- 由观察者独立对全部58个低密度病灶按预设顺序进行两次大小测量,采用组内相关系数(ICC)评估观察者间和观察者内一致性。
关键影像表现/征象
- 40-keV虚拟单能图像的CNR最高(4.3±2.6),显著高于标准重建(2.9±1.9;p<0.001)。
- 整体图像质量主观评分在两种重建技术间无显著差异(两者中位数均为4;p=0.147)。
- 病灶边缘勾画主观评分在40-keV虚拟单能图像更优(中位数5;p≤0.001)。
诊断标准与关键数值
- 40-keV虚拟单能图像病灶测量的观察者间ICC为0.98,标准重建为0.93。
- 40-keV虚拟单能图像病灶测量的观察者内ICC为0.99,标准图像系列为0.94。
临床应用与阅片要点
- 在脂肪肝患者中评估低密度肝转移瘤时,40-keV虚拟单能图像可能有助于改善病灶边缘勾画和对比显示。
- 两种重建方式下病灶大小测量的一致性均极好,测量结果稳定。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: To evaluate the impact of noise-optimized virtual monoenergetic imaging (VMI) on lesion demarcation and measuring accuracy of hypoattenuating liver metastases in patients with fatty liver disease compared to standard reconstructions.
MATERIALS AND METHODS: Twenty-eight patients (mean age 62.2 ± 7.7 years) with fatty liver disease and hypoattenuating liver metastases who underwent unenhanced and contrast-enhanced portal-venous dual-energy CT (DECT) were enrolled. Standard linearly blended and VMI series were reconstructed in 10-keV intervals. Lesion-to-parenchyma contrast-to-noise ratio (CNR) was calculated and the best VMI series was further investigated in a subjective evaluation of overall image quality and lesion demarcation. Size measurements were performed independently by measuring all hypodense lesions (n = 58) twice in a predefined sequence. Inter- and intra-rater agreement was assessed using intra-class correlation coefficient (ICC) statistics.
RESULTS: The calculated CNR was greatest at 40-keV VMI (4.3 ± 2.6), significantly higher compared to standard reconstructions (2.9 ± 1.9; p < 0.001). Subjective ratings for overall image quality showed no significant difference between the 2 reconstruction techniques (both medians 4; p = 0.147), while lesion margin demarcation was found to be superior for 40-keV VMI (median 5; p ≤ 0.001). Inter- (ICC, 0.98 for 40-keV VMI; ICC, 0.93 for standard reconstruction) and intra-rater (ICC, 0.99 for 40-keV VMI; ICC, 0.94 for standard image series) analysis showed an excellent agreement for lesion measurements in both reconstruction techniques.
CONCLUSION: Noise-optimized VMI reconstructions significantly improve contrast and lesion demarcation of hypoattenuating liver metastases in patients with the fatty liver disease compared to standard reconstruction.