文献速递介入技术与安全US

细针与粗针活检针填充造影剂六氟化硫:针体超声可见度的离体与体外评价

Filling of Fine and Core Biopsy Needles With the Contrast Agent Sulfur Hexafluoride: Ex Vivo and in vitro Evaluation of Ultrasound Needle Visibility.
PubMed 文献 · J Ultrasound Med · 2020-05-12

离体牛肝与体外水浴实验中,细针和粗针填充超声造影剂六氟化硫后,在造影特异性成像模式下评估针体可见度;粗针在造影模式下的主观可见度高于空气和生理盐水对照,细针与粗针的对比噪声比均升高,作者认为该方法有前景但需在体研究确认。

要点

  • 粗针填充六氟化硫后,在造影特异性成像模式下主观可见度高于空气和生理盐水对照。
  • 造影模式下细针与粗针的整段针体和针尖区对比噪声比均高于两种对照。
  • B 模式下填充造影剂未显示主观可见度改善,作者结论仅针对造影特异性成像模式。

核心结论

  • 针内填充超声造影剂是提高粗针在造影特异性成像模式下可见度的有前景的新方法,该发现需在体研究确认后才能评估其临床价值。

研究设计与样本

  • 离体与体外实验:实验 I 在屠宰牛肝中穿刺,实验 II 在水浴中穿刺,细针与粗针各在 B 模式和造影特异性成像模式下记录,每组 10 次观察。
  • 对照为空气和生理盐水,每个实验总样本 n=120;实验 I 由 4 名超声专家用 0–10 整数任意量表主观评估相对针体可见度,实验 II 计算整段针体与针尖区的对比噪声比。

关键影像表现/征象

  • 实验 I 中仅粗针在造影特异性成像模式下的主观可见度较空气和生理盐水两种对照升高(P < .01)。
  • 实验 II 中细针与粗针在造影特异性成像模式下,整段针体与针尖区的对比噪声比均较两种对照升高(P < .05)。

临床应用与阅片要点

  • 作者认为针内填充超声造影剂是提高粗针在造影特异性成像模式下可见度的有前景的新方法。
  • 作者指出该发现需在体研究确认后才能评估其临床价值。

局限性

  • 作者指出该发现需在体研究确认后才能评估其临床价值。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To investigate whether the ex vivo and in vitro ultrasound visibility of fine needles (FNs) and core biopsy needles (CNBs) can be improved by filling them with an ultrasound contrast agent.

METHODS: After needle filling with the contrast agent sulfur hexafluoride, punctures with FNs and CBNs were recorded in the B-mode and contrast-specific imaging mode (10 observations in each of the 4 groups). Recordings were made in both butchered bovine liver (experiment I) and a water bath (experiment II). Air and normal saline were used as controls (total n = 120 for each experiment). In experiment I, 4 ultrasound specialists subjectively assessed the relative needle visibility in the recordings by using an arbitrary scale (integers 0-10). In experiment II, the contrast-to-noise ratio was calculated for both the entire needle course and the needle tip area.

RESULTS: In experiment I, subjective visibility was increased compared with both controls only for CBNs in the contrast-specific imaging mode (P < .01). In experiment II, the contrast-to-noise ratio for both the entire needle course and the needle tip area increased compared with both controls for both FNs and CBNs in the contrast-specific imaging mode (P < .05).

CONCLUSIONS: Ultrasound contrast agent needle filling is a promising new method to increase the visibility of CBNs in the contrast-specific imaging mode. This finding needs to be confirmed in vivo before its clinical value can be assessed.

来源:J Ultrasound Med,查看原文。版权归原作者所有。

引用

  1. Thunswärd P, Nilsson A, Ahlström H, et al. Filling of Fine and Core Biopsy Needles With the Contrast Agent Sulfur Hexafluoride: Ex Vivo and in vitro Evaluation of Ultrasound Needle Visibility. J Ultrasound Med. 2020. PMID: 32395841. doi:10.1002/jum.15321

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。