大型血管外血肿的硬度对其体外沸腾组织粉碎液化敏感性的影响
体外研究用人和牛全血制备55–200 mL大血凝块,经3 h、3 d或8 d孵育回缩,用压痕仪和剪切波弹性成像测剪切模量,并以1.5 MHz聚焦超声30秒标准沸腾组织粉碎(BH)曝光产生的液化灶大小作为敏感性指标;人自然凝血块剪切模量0.52±0.08 kPa且不随孵育时间变化,显著低于牛血凝块2.85±0.17 kPa;在0.4–1.6 kPa范围内液化灶宽度比长度更受硬度影响,但所有血肿液化灶均大于相同参数下任何软组织,提示血肿对BH损伤敏感性更高。
要点
- 人自然凝血块剪切模量0.52±0.08 kPa,牛血凝块2.85±0.17 kPa,前者显著更低。
- 血肿液化灶宽度比长度更受硬度变化影响,但所有血肿灶均大于相同参数下软组织。
- 加凝血酶的牛抗凝血凝块可作为急性和慢性人血肿的体外模型。
诊断标准与关键数值
- 人自然凝血块剪切模量0.52±0.08 kPa,回缩程度(排出液体与原体积比)50%–58%,在0–8 d孵育时间内不随孵育时长变化。
- 牛血凝块剪切模量2.85±0.17 kPa,回缩程度5%–38%,显著高于人自然凝血块。
- 抗凝牛血制备的血凝块中,氯化钙浓度在5–40 mmol/L范围内变化不改变硬度;较低浓度和加入凝血酶使血凝块显著变软,与人自然凝血块相似。
- 在可达剪切模量范围0.4–1.6 kPa内,BH液化灶宽度比长度更受硬度变化影响;所有血肿液化灶均大于相同BH参数下任何软组织的液化灶。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Large intra-abdominal, retroperitoneal and intramuscular hematomas are common consequences of sharp and blunt trauma and post-surgical bleeds, and often threaten organ failure, compartment syndrome or spontaneous infection. Current therapy options include surgical evacuation and placement of indwelling drains that are not effective because of the viscosity of the organized hematoma. We have previously reported the feasibility of using boiling histotripsy (BH)-a pulsed high-intensity focused ultrasound method-for liquefaction of large volumes of freshly coagulated blood and subsequent fine-needle aspiration. The goal of this work was to evaluate the changes in stiffness of large coagulated blood volumes with aging and retraction in vitro, and to correlate these changes with the size of the BH void and, therefore, the susceptibility of the material to BH liquefaction. Large-volume (55-200 mL) whole-blood clots were fabricated in plastic molds from human and bovine blood, either by natural clotting or by recalcification of anticoagulated blood, with or without addition of thrombin. Retraction of the clots was achieved by incubation for 3 h, 3 d or 8 d. The shear modulus of the samples was measured with a custom-built indentometer and shear wave elasticity (SWE) imaging. Sizes of single liquefied lesions produced with a 1.5-MHz high-intensity focused ultrasound transducer within a 30-s standard BH exposure served as the metric for susceptibility of clot material to this treatment. Neither the shear moduli of naturally clotted human samples (0.52 ± 0.08 kPa), nor their degree of retraction (ratio of expelled fluid to original volume 50%-58%) depended on the length of incubation within 0-8 d, and were significantly lower than those of bovine samples (2.85 ± 0.17 kPa, retraction 5%-38%). In clots made from anticoagulated bovine blood, the variation of calcium chloride concentration within 5-40 mmol/L did not change the stiffness, whereas lower concentrations and the addition of thrombin resulted in significantly softer clots, similar to naturally clotted human samples. Within the achievable shear modulus range (0.4-1.6 kPa), the width of the BH-liquefied lesion was more affected by the changes in stiffness than the length of the lesion. In all cases, however, the lesions were larger compared with any soft tissue liquefied with the same BH parameters, indicating higher susceptibility of hematomas to BH damage. These results suggest that clotted bovine blood with added thrombin is an acceptable in vitro model of both acute and chronic human hematomas for assessing the efficiency of BH liquefaction strategies.
引用
- Khokhlova TD, Kucewicz JC, Ponomarchuk EM, et al. Effect of Stiffness of Large Extravascular Hematomas on Their Susceptibility to Boiling Histotripsy Liquefaction in Vitro. Ultrasound Med Biol. 2020. PMID: 32444137. doi:10.1016/j.ultrasmedbio.2020.04.023
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。