心脏计算机断层扫描心肌细胞外容积定量:双能碘法与标准减影法的比较

Myocardial Extracellular Volume Quantification Using Cardiac Computed Tomography: A Comparison of the Dual-energy Iodine Method and the Standard Subtraction Method.
PubMed 文献 · Acad Radiol · 2020-05-10

本研究以心脏磁共振为参考标准,回顾性分析21例患者平衡期心脏双层光谱探测器CT图像,比较标准减影法与双能碘法测量心肌细胞外容积的准确性。双能碘法与磁共振的相关性更高、95%一致性界限更窄、均方根误差更低,平均残差值显著低于标准减影法。结论认为双能碘法较标准减影法更准确,可获得与磁共振相当的细胞外容积值。

要点

  • 双能碘法与磁共振细胞外容积在室间隔和整体测量的相关性分别为r=0.94和r=0.95(p<0.01)。
  • 双能碘法与磁共振的95%一致性界限在室间隔和整体测量中均低于标准减影法。
  • 双能碘法的均方根误差低于标准减影法,平均残差值显著低于标准减影法。

核心结论

  • 双能碘法所得心肌细胞外容积定量较标准减影法更准确,且与心脏磁共振所得细胞外容积值相当。

研究设计与样本

  • 回顾性分析21例患者,采用双层光谱探测器CT采集平衡期心脏图像并以心脏磁共振为参考标准。
  • 分别用标准减影法(ECVsub)和双能碘法(ECViodine)计算CT细胞外容积,评估各方法与磁共振细胞外容积的相关性、一致性及误差。

诊断标准与关键数值

  • 室间隔与整体测量中,ECVsub与ECViodine相关性分别为r=0.95(p<0.01)和r=0.91(p<0.01)。
  • ECVsub与磁共振细胞外容积相关性为室间隔r=0.90(p<0.01)、整体r=0.84(p<0.01)。
  • ECViodine与磁共振细胞外容积相关性为室间隔r=0.94(p<0.01)、整体r=0.95(p<0.01)。
  • Bland-Altman分析显示ECViodine与磁共振的95%一致性界限在室间隔和整体测量中均低于ECVsub与磁共振的95%一致性界限。
  • ECVsub的均方根误差高于ECViodine,ECVsub的平均残差值显著高于ECViodine。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: To clarify the accuracy of two measurement methods for myocardial extracellular volume (ECV) quantification (ie, the standard subtraction method [ECVsub] and the dual-energy iodine method [ECViodine]) with the use of cardiac CT in comparison to cardiac magnetic resonance imaging (CMR) as a reference standard.

MATERIALS AND METHODS: Equilibrium phase cardiac images of 21 patients were acquired with a dual-layer spectral detector CT and CMR, and the images were retrospectively analyzed. CT-ECV was calculated using ECVsub and ECViodine. The correlation between the ECV values measured by each method was assessed. Bland-Altman analysis was used to identify systematic errors and to determine the limits of agreement between the CT-ECV and CMR-ECV values. Root mean squared errors and residual values for the ECVsub and ECViodine were also assessed.

RESULTS: The correlations between ECVsub and ECViodine for both septal and global measurement were r = 0.95 (p < 0.01) and 0.91 (p < 0.01), respectively, while those between the mean ECVsub and CMR-ECV were r = 0.90 (septal, p < 0.01) and 0.84 (global, p < 0.01), and those between ECViodine and CMR-ECV were r = 0.94 (septal, p < 0.01) and 0.95 (global, p < 0.01). Bland-Altman plots showed lower 95% limits of agreement between ECViodine and CMR-ECV compared with that between ECVsub and CMR-ECV in both septal and global measurement. The root mean squared error of ECVsub was higher than that of ECViodine. The mean residual value of ECVsub was significantly higher than that of ECViodine.

CONCLUSION: ECViodine yielded more accurate myocardial ECV quantification than ECVsub, and provided a comparable ECV value to that obtained by CMR.

来源:Acad Radiol,查看原文。版权归原作者所有。