用于临床与研究应用的心脏组织模拟弹道凝胶超声模体

Cardiac Tissue-Mimicking Ballistic Gel Phantom for Ultrasound Imaging in Clinical and Research Applications.
PubMed 文献 · Ultrasound Med Biol · 2020-05-16

研究评估弹道凝胶作为组织模拟材料在人体心脏模体中的超声成像性能,其声速为 1537 ± 39 m/s,衰减系数为 1.07 dB/cm·MHz,均在心脏组织已报道范围内;基于人体解剖分割和常规塑料成型技术制作的心脏模型可在心腔内超声系统中识别关键解剖特征,材料耐用且可低成本重复生产。

要点

  • 弹道凝胶声速 1537 ± 39 m/s,接近心脏组织典型值约 1576 m/s。
  • 衰减系数 1.07 dB/cm·MHz,位于心脏组织已报道范围 0.81–1.81 dB/cm·MHz 内。
  • 基于人体解剖分割和常规塑料成型技术可制作可重复、低成本的全心脏复杂模体。

关键影像表现/征象

  • 在 2–12 MHz 范围内测得声速为 1537 ± 39 m/s,接近心脏组织典型值约 1576 m/s。
  • 衰减系数为 1.07 dB/cm·MHz,位于心脏组织已报道范围 0.81–1.81 dB/cm·MHz 内。
  • 基于人体解剖的心脏模型使用心腔内超声成像系统可观察、捕获并识别关键解剖特征。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Ballistic gel was investigated as a tissue-mimicking material in an anthropomorphic cardiac phantom for ultrasound imaging. The gel was tested for its acoustic properties and its compatibility with conventional plastics molding techniques. Speed of sound and attenuation were evaluated in the range 2-12 MHz. The speed of sound was 1537 ± 39 m/s, close to typical values for cardiac tissue (∼1576 m/s). The attenuation coefficient was 1.07 dB/cm·MHz, within the range of values previously reported for cardiac tissue (0.81-1.81 dB/cm·MHz). A cardiac model based on human anatomy was developed using established image segmentation processes and conventional plastic molding techniques. Key anatomic features were observed, captured and identified in the model using an intracardiac ultrasound imaging system. These favorable results along with the material's durability and processes that allow for repetitive production of detailed whole-heart models at low cost are promising. There are numerous applications for geometrically complex phantoms in research, training, device development and clinical use.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。