CT引导介入操作的国家剂量参考水平——一项提案

National dose reference levels in computed tomography-guided interventional procedures-a proposal.
PubMed 文献 · Eur Radiol · 2020-05-02

为落实2013/59/Euratom欧洲指令,该研究在49个中心收集5001例CT引导下介入操作,涵盖胸部、腹盆腔及骨关节共17类操作,以第3四分位数作为参考水平、中位数作为优化目标值,提出按总剂量长度乘积和螺旋采集次数计的参考水平,并发现同一操作在不同中心的患者剂量差异很大。

要点

  • 17类CT引导介入操作的剂量参考水平以总剂量长度乘积第3四分位数表示,螺旋采集次数参考水平同时给出。
  • 椎体或椎旁浸润参考水平为375 mGy.cm、2次螺旋采集,椎体成形术为1630 mGy.cm、3次,活检为845 mGy.cm、4次,肿瘤消融为1950 mGy.cm、8次,引流为1090 mGy.cm、5次。
  • 总剂量长度乘积和螺旋采集次数随操作复杂度增加,同一类操作在不同中心的患者剂量差异很大。

核心结论

  • 这是首个提出CT引导下介入操作国家参考水平的多中心全国性调查,所提参考水平可用于影像科室对标其他机构并优化扫描方案。

研究设计与样本

  • 全国性多中心调查,纳入49个中心、5001例操作,覆盖胸部、腹盆腔及骨关节17类介入操作(经皮浸润、椎体成形术、活检、引流、肿瘤消融)。
  • 每中心每类操作记录10至20例患者的总剂量长度乘积、螺旋采集次数,以及螺旋、序列或透视采集的总剂量长度乘积。
  • 参考水平取分布的第3四分位数,优化过程目标值取中位数,并与既往发表研究比较。

诊断标准与关键数值

  • 椎体或椎旁浸润参考水平:总剂量长度乘积375 mGy.cm,螺旋采集次数2次。
  • 椎体成形术参考水平:总剂量长度乘积1630 mGy.cm,螺旋采集次数3次。
  • 活检参考水平:总剂量长度乘积845 mGy.cm,螺旋采集次数4次。
  • 肿瘤消融参考水平:总剂量长度乘积1950 mGy.cm,螺旋采集次数8次。
  • 引流参考水平:总剂量长度乘积1090 mGy.cm,螺旋采集次数5次。

临床应用与阅片要点

  • 影像科室可用所提参考水平对标本机构同类CT引导介入操作的总剂量长度乘积与螺旋采集次数。
  • 参考水平以第3四分位数定义、优化目标值以中位数定义,可用于优化本机构扫描方案。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To establish national reference levels (RLs) in interventional procedures under CT guidance as required by the 2013/59/Euratom European Directive.

METHODS: Seventeen categories of interventional procedures in thoracic, abdominopelvic, and osteoarticular specialties (percutaneous infiltration, vertebroplasty, biopsy, drainage, tumor destruction) were analyzed. Total dose length product (DLP), number of helical acquisitions (NH), and total DLP for helical, sequential, or fluoroscopic acquisitions were recorded for 10 to 20 patients per procedure at each center. RLs were calculated as the 3rd quartiles of the distributions and target values for optimization process (TVOs) as the median. RLs and TVOs were compared with previously published studies.

RESULTS: Results on 5001 procedures from 49 centers confirmed the great variability in patient dose for the same category of procedures. RLs were proposed for the DLPs and NHs in the seventeen categories. RLs in terms of DLP and NH were 375 mGy.cm and 2 NH for spinal or peri-spinal infiltration, 1630 mGy.cm and 3 NH for vertebroplasty, 845 mGy.cm and 4 NH for biopsy, 1950 mGy.cm and 8 NH for destruction of tumors, and 1090 mGy.cm and 5 NH for drainage. DLP and NH increased with the complexity of procedures.

CONCLUSIONS: This study was the first nationwide multicentric survey to propose RLs for interventional procedures under CT guidance. Heterogeneity of practice in centers were found with different levels of patient doses for the same procedure. The proposed RLs will allow imaging departments to benchmark their practice with others and optimize their protocols.

KEY POINTS: • National reference levels are proposed for 17 categories of interventional procedures under CT guidance. • Reference levels are useful for benchmarking practices and optimizing protocols. • Reference levels are proposed for dose length product and the number of helical acquisitions.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Greffier J, Ferretti G, Rousseau J, et al. National dose reference levels in computed tomography-guided interventional procedures-a proposal. Eur Radiol. 2020. PMID: 32361774. doi:10.1007/s00330-020-06903-9

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。