文献速递神经技术与安全MR

评估使用压缩SENSE实现全脑覆盖的高分辨率T1灌注MRI的可行性:应用于胶质瘤分级

Evaluating feasibility of high resolution T1-perfusion MRI with whole brain coverage using compressed SENSE: Application to glioma grading.
PubMed 文献 · Eur J Radiol · 2020-05-11

本研究纳入5名健康志愿者、1个体模和126例脑肿瘤患者的MRI数据,评估使用压缩SENSE(CSENSE)优化的全脑覆盖高分辨率T1灌注MRI方案(方案2)与本单位常规部分脑覆盖方案(方案1)在区分高级别胶质瘤(HGG)和低级别胶质瘤(LGG)中的效能。结果显示结构图像在加速因子R=4时均达诊断质量,R=6时出现伪影;两方案计算的肿瘤组织所有参数和健康灰质血流动力学参数无显著差异;方案2参数在胶质瘤分级中表现更优,两方案均以rCBF表现最差(AUC分别为0.759和0.851),所有参数联合表现最佳(AUC分别为0.890和0.964)。结论认为CSENSE(R=4)可用于改善T1灌注分析的分辨率和脑覆盖以区分胶质瘤。

要点

  • CSENSE加速因子R=4时结构图像均达诊断质量,R=6时出现伪影。
  • 两方案计算的肿瘤组织所有参数及健康灰质血流动力学参数无显著差异。
  • 方案2参数在胶质瘤分级中表现更优,所有参数联合AUC达0.964。

关键影像表现/征象

  • 结构图像在加速因子R=4时均达诊断质量,R=6时结构图像和参数图出现伪影。
  • 健康T1加权FFE图像的NMSE和体模图像的COV随R增加而增高,图像在R=4时提供最佳质量。

诊断标准与关键数值

  • 方案1中rCBF的AUC为0.759,方案2中rCBF的AUC为0.851。
  • 方案1中所有参数联合的AUC为0.890,方案2中所有参数联合的AUC为0.964。
  • 两方案计算的肿瘤组织所有参数和健康灰质血流动力学参数无显著差异。

临床应用与阅片要点

  • 使用CSENSE(R=4)可在保持诊断质量的同时改善T1灌注分析的脑覆盖和分辨率。
  • 方案2参数在胶质瘤分级中表现优于方案1,所有参数联合诊断效能最佳。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To evaluate the efficacy of optimized T1-Perfusion MRI protocol (protocol-2) with whole brain coverage and improved spatial resolution using Compressed-SENSE (CSENSE) to differentiate high-grade-glioma (HGG) and low-grade-glioma (LGG) and to compare it with the conventional protocol (protocol-1) with partial brain coverage used in our center.

METHODS: This study included MRI data from 5 healthy volunteers, a phantom and 126 brain tumor patients. Current study had two parts: To analyze the effect of CSENSE on 3D-T1-weighted (W) fast-field-echo (FFE) images, T1-W, dual-PDT2-W turbo-spin-echo images and T1 maps, and to evaluate the performance of high resolution T1-Perfusion MRI protocol with whole brain coverage optimized using CSENSE. Coefficient-of-Variation (COV), Relative-Percentage-Error (RPE), Normalized-Mean-Squared-Error (NMSE) and qualitative scoring were used for the former study. Tracer-kinetic (Ktrans,ve,vp) and hemodynamic (rCBV,rCBF) parameters computed from both protocols were used to differentiate LGG and HGG.

RESULTS: The image quality of all structural images was found to be of diagnostic quality till R = 4. NMSE in healthy T1-W-FFE images and COV in phantom images increased with-respect-to R and images provided optimum quality till R = 4. Structural images and maps exhibited artefacts from R = 6. All parameters in tumor tissue and hemodynamic parameters in healthy gray matter tissue computed from both protocols were not significantly different. Parameters computed from protocol-2 performed better in terms of glioma grading. For both protocols, rCBF performed least (AUC = 0.759 and 0.851) and combination of all parameters performed best (AUC = 0.890 and 0.964).

CONCLUSION: CSENSE (R = 4) can be used to improve the resolution and brain coverage for T1-Perfusion analysis used to differentiate gliomas.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Sasi S D, Ramaniharan AK, Bhattacharjee R, et al. Evaluating feasibility of high resolution T1-perfusion MRI with whole brain coverage using compressed SENSE: Application to glioma grading. Eur J Radiol. 2020. PMID: 32464580. doi:10.1016/j.ejrad.2020.109049

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。