采用第三代重建算法的脑CT血管成像在降低对比剂和辐射剂量的同时改善图像质量
回顾性纳入100例连续患者(低剂量组与常规剂量组各50例),比较第三代重建算法联合低对比剂剂量-低kVp技术(LD-CTA)与常规对比剂剂量120 kVp联合正弦图确认迭代重建算法(ND-CTA)的脑CTA图像质量。LD-CTA组主观图像质量相关系数良好,双侧大脑中动脉及基底动脉中段信噪比和对比噪声比均更优(p<0.0001),且对比剂用量更低(49 cc vs 97 cc)、辐射暴露更低(DLP/CTDI两组268.3/80.7 vs 519.5/36.08,均<0.0001)。
要点
- LD-CTA组对比剂用量49 cc,低于ND-CTA组的97 cc。
- LD-CTA组DLP/CTDI为268.3/80.7,低于ND-CTA组的519.5/36.08(均<0.0001)。
- 双侧大脑中动脉及基底动脉中段SNR和CNR在LD-CTA组更优(p<0.0001)。
关键影像表现/征象
- LD-CTA组双侧大脑中动脉及基底动脉中段SNR和CNR均优于ND-CTA组(p<0.0001)。
- 两名阅片者主观图像质量的平均一致率:第1组为95.2%,第2组为89.2%。
诊断标准与关键数值
- 主观图像质量采用类Likert量表对多个血管节段评分,两名阅片者间ICC为0.51-0.63,呈中度相关。
- SNR和CNR在双侧大脑中动脉M1段中点和基底动脉中段测量。
临床应用与阅片要点
- LD-CTA组对比剂用量为49 cc,ND-CTA组为97 cc。
- LD-CTA组辐射暴露DLP/CTDI为268.3/80.7,ND-CTA组为519.5/36.08,两组比较均<0.0001。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: We hypothesized that cerebral CT angiogram performed using third-generation reconstruction algorithm and lower contrast dose-low-kVp technique (LD-CTA) will provide better image quality when compared with regular contrast dose CTA at 120 kVp using a sinogram-affirmed iterative reconstruction algorithm (ND-CTA).
METHODS: Retrospective imaging review of 100 consecutive patients (50 each in LD- and ND-CTA groups). Two readers independently assessed the subjective image quality across multiple vascular segments on a Likert-like scale. Differences in contrast dose, CT dose index (CTDI), and dose length product (DLP) were compared using Mann-Whitney U test. Fisher's exact test was used to compare subjective image quality. Similarly, contrast- and signal-to-noise ratios (CNR and SNR) were compared in the mid-M1 MCA vessels bilaterally and the mid-basilar artery using Mann-Whitney U test. Interclass correlation coefficient (ICC) was calculated for the SNR/CNR values.
RESULTS: Both observers showed excellent correlation in subjective image quality (mean percentage agreement of 95.2% for group 1 versus 89.2% for group 2). LD-CTA group showed better SNR and CNR (p < 0.0001) for both MCA vessels and the mid-basilar artery. Interclass correlation coefficient showed moderate correlation (0.51-0.63) between readers. LD-CTA group also used lower contrast (49 cc versus 97 cc in ND-CTA) and had lower radiation exposure (DLP/CTDI for both groups 268.3/80.7 vs 519.5/36.08, both < 0.0001).
CONCLUSION: Next-generation reconstruction algorithm and low-kV scanning significantly improved image quality on cerebral CTA images despite lower contrast dose and, in addition, have lower radiation exposure.
引用
- Bathla G, Priya S, Samaniego E, et al. Cerebral computed tomographic angiography using third-generation reconstruction algorithm provides improved image quality with lower contrast and radiation dose. Neuroradiology. 2020. PMID: 32277245. doi:10.1007/s00234-020-02406-y
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。