超声探头及多用途超声耦合剂瓶口在无菌清洁技术前后的细菌污染鉴定与分析

Identification and Analysis of Bacterial Contamination of Ultrasound Transducers and Multiuse Ultrasound Transmission Gel Bottle Tips Before and After the Aseptic Cleansing Technique.
PubMed 文献 · J Ultrasound Med · 2020-04-27

本研究对一个主要临床机构内18个临床护理点的超声探头和多用途耦合剂瓶口进行前瞻性盲法描述性分析,比较无菌清洁前后的细菌培养结果。清洁前73.8%的样本培养出临床相关细菌,清洁后降至16.3%,临床相关细菌生长减少78.0%(似然比57.10,P < .001)。结论支持美国超声医学研究所2018年指南,强调根据检查类型进行适当探头准备。

要点

  • 清洁前73.8%的探头样本培养出临床相关细菌,清洁后降至16.3%,减少78.0%。
  • 模拟葡萄球菌占所有阳性培养样本的31.0%。
  • 无菌清洁可有效清除大部分细菌负荷,但部分细菌持续存在,超声引导介入操作存在院内感染风险。

核心结论

  • 超声探头具有明显的临床相关细菌负荷,可能成为潜在感染媒介;无菌清洁能有效清除大部分细菌负荷,但部分细菌持续存在,超声引导介入操作存在院内感染风险;研究结果支持美国超声医学研究所2018年指南。

研究设计与样本

  • 前瞻性盲法描述性分析研究,纳入单一主要临床机构内18个不同临床护理点,共分析188份样本:80份来自超声探头清洁前后,13份来自多用途耦合剂瓶口清洁前后,2份为数据收集者笔和胸牌清洁前样本;采用胰蛋白酶大豆琼脂接触平板、盲法培养、革兰染色及微测序进行菌种鉴定。

诊断标准与关键数值

  • 清洁前59份样本(73.8%)培养出临床相关细菌,21份样本(26.3%)未培养出。
  • 清洁后13份样本(16.3%)培养出临床相关细菌,临床相关细菌生长减少78.0%(似然比57.10;P < .001)。
  • 模拟葡萄球菌占所有阳性培养样本的31.0%。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To provide a descriptive analysis for species identification of culture and Gram stain results from ultrasound transducers and multiuse ultrasound transmission gel bottle tips in active clinical use and to compare bacterial cultures from ultrasound transducers before and after aseptic cleansing.

METHODS: A prospective blinded descriptive analytic study of 18 distinct clinical care sites within a single primary clinical institution was conducted. Before and after a disinfectant towel cleanse, transducers were pressed against tryptic soy agar contact plates. Plates were deidentified and submitted for blind incubation, Gram staining, and species identification with microsequencing. Results were classified as clinically relevant (CR) or non-clinically relevant. In total, 188 samples were analyzed: 80 from ultrasound transducers before and cleansing, 13 from multiuse gel bottle tips before and after cleansing, and 2 precleansing samples from the data collector's pen and badge.

RESULTS: Fifty-nine precleansing samples (73.8%) grew cultures with CR bacteria, and 21 samples (26.3%) did not. Staphylococcus simulans represented 31.0% of all positive culture samples. Thirteen postcleansing samples (16.3%) grew cultures with CR bacteria, equating to a 78.0% reduction of CR bacterial growth (likelihood ratio, 57.10; P < .001).

CONCLUSIONS: Ultrasound transducers have a notable CR bacterial burden and may serve as potential infective vectors. Aseptic cleansing effectively eliminates most of the bacterial load from ultrasound transducers, but some bacteria persist, presenting a risk of nosocomial infection with ultrasound-guided interventions. These findings support American Institute of Ultrasound in Medicine 2018 guidelines intended to ensure an appropriate level of transducer preparation based on the examination type while emphasizing rational infection control measures to minimize the risk of potential patient harm.