商用介入透视系统在透视引导介入操作中的辐射剂量估算与图像质量比较研究
采用体模研究和临床操作比较两种商用单平面平板探测器介入透视系统的辐射剂量与图像质量。体模研究显示Artis Q在模拟患者厚度30-40 cm时空气比释动能率显著更高;4381例临床资料显示AlluraClarityIQ的累积空气比释动能和剂量面积乘积在部分操作中显著更低。40例同一医师再次栓塞的配对分析中AlluraClarityIQ的累积空气比释动能和剂量面积乘积分别降低45%和31%,主观图像质量评分更高。
要点
- 体模研究显示模拟患者厚度30-40 cm时Artis Q的空气比释动能率显著高于AlluraClarityIQ(p<0.05)。
- 4381例临床资料显示AlluraClarityIQ的累积空气比释动能中位数低29%-58%。
- 40例配对栓塞操作中AlluraClarityIQ的累积空气比释动能和剂量面积乘积分别降低45%和31%。
诊断标准与关键数值
- 图像质量由两名盲法介入放射科医师采用五点量表(1-5分,5分最佳)评价五项参数。
- 体模研究中模拟患者厚度为20-40 cm,系统报告的空气比释动能率在30-40 cm时Artis Q显著更高(p<0.05)。
临床应用与阅片要点
- AlluraClarityIQ在部分操作中累积空气比释动能中位数低29%-58%,配对栓塞操作中累积空气比释动能和剂量面积乘积分别降低45%和31%。
- AlluraClarityIQ五项图像质量参数评分平均差异为1.3-1.6,评分显著更高(p<0.005)。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: To investigate differences in radiation dose and image quality for single-plane flat-panel-detector based interventional fluoroscopy systems from two vendors using phantom study and clinical procedures.
MATERIALS AND METHODS: AlluraClarityIQ (Philips) and Artis Q (Siemens-Healthineers) interventional fluoroscopy systems were evaluated. Phantom study included comparison of system-reported air-kerma rates (AKR) for clinical protocols with simulated patient thicknesses (20-40 cm). Differences in system-reported radiation dose estimates, cumulative-air-kerma (CAK) and kerma-area-product (KAP), for different clinical procedures were investigated. Subset analysis investigated differences in CAK, KAP and other factors affecting radiation dose when the same patients underwent repeat embolization procedures performed by the same physician on the two different fluoroscopy systems. Two blinded interventional radiologists reviewed image-quality for these procedures using a five-point scale (1-5; 5-best) for five parameters.
RESULTS: Phantom study revealed that air-kerma rates was significantly higher for Artis Q system for 30-40cm of simulated patient thicknesses (p < 0.05). Overall data analysis from 4381 clinical cases revealed significant differences in CAK and KAP for certain procedures (p < 0.05); with significantly lower values for AlluraClarityIQ systems (median CAK lower by: 29%-58%). Subset analysis with 40 patients undergoing repeat embolization procedures on both systems revealed that median CAK and KAP were significantly lower for AlluraClarityIQ systems (p < 0.02) by 45% and 31%, respectively. Image quality scores for AlluraClarityIQ systems were significantly greater (mean difference range for five parameters: 1.3-1.6; p < 0.005).
CONCLUSION: Radiation dose and image quality differences were observed between AlluraClarityIQ and Artis Q systems. AlluraClarityIQ systems showed lower radiation utilization and an increase in subjective perception of image quality.
引用
- Trunz LM, Eschelman DJ, Gonsalves CF, et al. Investigation of Radiation Dose Estimates and Image Quality Between Commercially Available Interventional Fluoroscopy Systems for Fluoroscopically Guided Interventional Procedures. Acad Radiol. 2020. PMID: 32224035. doi:10.1016/j.acra.2020.02.024
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。