骨选择性MRI作为颅面成像中CT的无辐射替代方案
本研究评估一种新型骨选择性MRI技术用于生成三维人体颅骨渲染图的可行性;对一具尸体颅骨和五名健康成人进行双射频脉冲双回波三维超短回波时间序列扫描,每例扫描在6分钟内完成,并与薄层头部CT比较。基于MR的测量与基于CT的测量平均百分比差异为2.3%-5.0%,Lin一致性相关系数为0.998-1.000,提示该方法与CT相比具有良好可靠性。
要点
- 双射频双回波三维超短回波时间序列可在6分钟内完成扫描并清晰显示颅面骨骼结构。
- 基于MR的三维颅骨渲染图测量与CT测量平均百分比差异为2.3%-5.0%,一致性相关系数为0.998-1.000。
- 三维渲染图中部分体素被错误纳入或排除,目前受分割校正所需时间和手动输入限制。
关键影像表现/征象
- 基于MR的三维渲染图描绘了大多数颅面特征(例如颧弓),但部分体素在渲染图中被错误纳入或排除。
诊断标准与关键数值
- 基于MR的测量与基于CT的测量平均百分比差异范围为2.3%-5.0%。
- 基于MR与基于CT测量的Lin一致性相关系数范围为0.998-1.000。
临床应用与阅片要点
- 作者认为该技术可在临床可行的成像时间内产生高分辨率骨特异性图像,清晰显示颅面骨骼结构,一致性系数提示该方法与CT相比具有良好可靠性。
- 作者指出该方法目前受分割校正所需时间和手动输入限制,需进一步研究以获得更准确的三维渲染图和用于儿科患者扫描。
局限性
- 三维渲染图中部分体素被错误纳入或排除。
- 该方法目前受分割校正所需时间和手动输入限制。
- 需进一步研究以获得更准确的三维渲染图和用于儿科患者扫描。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONAL AND OBJECTIVES: Computed tomography (CT) is the clinical gold-standard for high-resolution 3D visualization of cortical bone structures. However, ionizing radiation is of concern, particularly for pediatric patients. This study evaluates the feasibility of producing 3D human skull renderings using a novel bone-selective magnetic resonance imaging technique.
MATERIALS AND METHODS: A dual-radiofrequency pulse, dual-echo, 3D ultrashort echo time sequence was applied for scanning of a cadaver skull and five healthy adult subjects. Scans were each completed within 6 minutes. Semiautomatic segmentation of bone voxels was performed using ITK-SNAP software, leading to 3D renderings of the skulls. For comparison, thin-slice head CT scans were performed. Mimics software was used to measure eight anatomic distances from 3D renderings. Lin's Concordance Correlation test was applied to assess agreement between measurements from MR-based and CT-based 3D skull renderings.
RESULTS: The 3D rendered MR images depict most craniofacial features (e.g., zygomatic arch), although some voxels were erroneously included or excluded in the renderings. MR-based measurements differed from CT-based measurements by mean percent difference ranging from 2.3%-5.0%. Lin's Concordance Correlation Coefficients for MR-based vs CT-based measurements ranged from 0.998-1.000.
CONCLUSION: The proposed dual-radiofrequency dual-echo 3D ultrashort echo time imaging technique produces high-resolution bone-specific images within a clinically feasible imaging time, leading to clear visualization of craniofacial skeletal structures. Concordance coefficients suggest good reliability of the method compared to CT. The method is currently limited by time and manual input necessary for segmentation correction. Further investigation is needed for more accurate 3D renderings and for scanning of pediatric patients.
引用
- Zhang R, Lee H, Zhao X, et al. Bone-Selective MRI as a Nonradiative Alternative to CT for Craniofacial Imaging. Acad Radiol. 2020. PMID: 32299762. doi:10.1016/j.acra.2020.03.001
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。