文献速递骨肌DR

EOS与传统X线评估成人下肢畸形的可靠性比较

Reliability of EOS compared to conventional radiographs for evaluation of lower extremity deformity in adult patients.
PubMed 文献 · Skeletal Radiol · 2020-04-15

前瞻性纳入10例(20条下肢)平均年龄31.6岁的创伤后畸形患者,由两名独立观察者间隔2周分别在传统X线与二维EOS上测量下肢全长正位参数。EOS与X线在股骨长度、胫骨长度和下肢总长度上差异有统计学意义但临床意义不大,所有测量一致性均高(r=0.84-0.99)。结论认为二维EOS用于下肢测量可重复、可靠,与站立位下肢全长X线金标准相当。

要点

  • EOS与站立位下肢全长X线各测量参数一致性高(r=0.84-0.99)。
  • EOS测得的股骨长度和下肢总长度均小于传统X线,差异有统计学意义但临床意义不大。
  • EOS测得的膝内外翻角度(V/V)多数较传统X线偏大。

核心结论

  • 二维EOS用于下肢测量可重复、可靠,与站立位下肢全长X线金标准相当。

研究设计与样本

  • 前瞻性研究,纳入10例(20条下肢)平均年龄31.6岁(范围21-39岁)的创伤后畸形患者,因评估截骨术和/或韧带与软骨重建而就诊。
  • 两名独立观察者间隔2周分别在传统X线和二维EOS图像上完成下肢全长正位测量。
  • 测量参数包括负重轴(WBA)、膝内外翻角度(V/V)、股骨长度(FL)、胫骨长度(TL)、股骨机械轴(FMA)、胫骨机械轴(TMA)和下肢总长度(TLL)。

关键影像表现/征象

  • EOS与X线在膝内外翻角度(V/V)、股骨长度(FL)和下肢总长度(TLL)上差异有统计学意义(均p<0.01)。
  • 相当比例患者的EOS膝内外翻角度(V/V)测量值持续大于传统X线。
  • EOS的股骨长度(FL)和下肢总长度(TLL)测量值持续小于传统X线。
  • EOS与X线测量值之间的差异虽有统计学意义,但数值差异小、无临床意义。

诊断标准与关键数值

  • 采用随机效应模型计算组内相关系数(ICC),设定为一致性。
  • 采用学生双样本t检验、配对双样本t检验和Pearson相关r统计一致性比较传统X线与EOS的观察者间效度。
  • 所有测量的一致性均保持高位(r=0.84-0.99)。

临床应用与阅片要点

  • 二维EOS测量下肢力线及长度参数可重复、可靠,与站立位下肢全长X线金标准相当,可用于日常下肢畸形评估。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The purpose of this study was to compare reliability of lower extremity imaging measurements using EOS and conventional X-ray (CR) of adult patients with mechanical axis malalignment.

MATERIALS AND METHODS: Ten patients (20 lower limbs) of mean age of 31.6 years (range 21-39) with post-traumatic deformities who presented for evaluation of osteotomies and/or ligament and cartilage reconstructions were prospectively enrolled. Two independent observers performed full-length anterior-posterior (AP) measurements 2 weeks apart on both CXR and two-dimensional (2D) EOS images. Measurements included weight-bearing axis (WBA), varus/valgus angle (V/V), femoral length (FL), tibial length (TL), femoral mechanical axis (FMA), tibial mechanical axis (TMA), and total limb length (TLL). Reliability was determined with random effects modeling of intraclass correlation coefficients (ICC) set to consistency. Three statistical operations were performed to compare interrater validity in CXR and EOS: students' two-sample t test, paired two-sample t test, and Pearson's correlative r-statistical agreement.

RESULTS: There was a statistically significant difference for V/V, FL, and TLL (all p < 0.01) between CXR and EOS. A relatively large proportion of the population consistently had larger V/V measures for EOS compared to CXR. In contrast, the FL and TLL measures were consistently larger for CXR compared to EOS. The differences between CXR and EOS measurements were statistically significant, though the small differences in values were not clinically meaningful. Agreement of all measures remained high (r = 0.84-0.99).

CONCLUSION: Using 2D EOS for lower extremity measurements is reproducible, reliable, and comparable to the gold standard, standing long leg radiographs.

来源:Skeletal Radiol,查看原文。版权归原作者所有。

引用

  1. Wise KL, Kelly BJ, Agel J, et al. Reliability of EOS compared to conventional radiographs for evaluation of lower extremity deformity in adult patients. Skeletal Radiol. 2020. PMID: 32291475. doi:10.1007/s00256-020-03425-9

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。