二维合成乳腺摄影与数字乳腺摄影在筛查中检测微钙化的性能比较
本研究为回顾性多阅片者配对病例研究,比较二维合成乳腺摄影(SM)与全视野数字乳腺摄影(FFDM)检测微钙化的性能。SM的AUC为91%(95%CI 83-97%),FFDM为88%(95%CI 79-95%),SM不劣于FFDM(p=0.011)。SM特异度和阳性预测值低于FFDM(91% vs 98%,p=0.034;87% vs 96%,p=0.034),但微钙化在SM上更明显(p<0.0001)。
要点
- SM检测微钙化的诊断准确率不劣于FFDM(AUC 91% vs 88%,非劣效性p=0.011)。
- SM特异度和阳性预测值低于FFDM(91% vs 98%;87% vs 96%,p=0.034)。
- 所有阅片者认为钙化在SM上更明显(p<0.0001),但诊断信心无显著差异。
关键影像表现/征象
- 所有阅片者认为钙化在SM上更明显(p<0.0001)。
- 两种模态的主观诊断信心无显著差异。
诊断标准与关键数值
- SM的AUC为91%(95%CI 83-97%),FFDM的AUC为88%(95%CI 79-95%),p=0.392,SM非劣效于FFDM(p=0.011)。
- SM与FFDM的敏感度分别为77% vs 73%(p=0.366),阴性预测值分别为84% vs 82%(p=0.598)。
- SM与FFDM的特异度分别为91% vs 98%(p=0.034),阳性预测值分别为87% vs 96%(p=0.034)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE. The purpose of this study is to compare the performance of 2D synthetic mammography (SM) to that of full-field digital mammography (FFDM) in the detection of microcalcifications and to evaluate radiologists' preference between the two imaging modalities for assessing calcifications. MATERIALS AND METHODS. A fully crossed, mode-balanced, paired-case (n = 160), retrospective, multireader (n = 3) performance study was implemented to compare screening mammograms acquired with digital breast tomosynthesis and both FFDM and SM between 2015 and 2017. The study cohort included 70 patients with mammograms recalled for microcalcifications (35 with malignant findings and 35 with benign findings) and was supplemented with 90 patients with mammograms with negative findings. In separate sessions, readers interpreted SM or FFDM images by recording a BI-RADS assessment and the probability of malignancy. In a final session that included 70 mammograms with microcalcifications, readers recorded their subjective assessment of microcalcification conspicuity and diagnostic confidence. RESULTS. There was no difference in diagnostic accuracy as assessed by comparing the likelihood of malignancy based on the AUC of plotted ROCs, with AUCs of 91% (95% CI, 83-97%) and 88% (95% CI, 79-95%) observed for SM and FFDM, respectively (p = 0.392), and with noninferiority of SM compared with FFDM (p = 0.011). No significant difference was observed between SM and FFDM in terms of sensitivity (77% vs 73%, respectively; p = 0.366) or negative predictive value (84% vs 82%, respectively; p = 0.598). The specificity and positive predictive value of SM were lower than those of FFDM (91% vs 98%, respectively [p = 0.034], and 87% vs 96%, respectively [p = 0.034]). All readers found calcifications to be more conspicuous on SM (p < 0.0001); however, no significant difference in subjective diagnostic confidence was seen. CONCLUSION. SM is noninferior to FFDM in the detection of microcalcifications. Despite the increased conspicuity of microcalcifications on SM, the subjective diagnostic confidence in the two modalities is comparable.
引用
- Dodelzon K, Simon K, Dou E, et al. Performance of 2D Synthetic Mammography Versus Digital Mammography in the Detection of Microcalcifications at Screening. AJR Am J Roentgenol. 2020. PMID: 32255687. doi:10.2214/AJR.19.21598
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。