欧洲指令2013/59/Euratom转化时代乳腺X线摄影剂量学的准确性
本研究纳入4028例接受五种不同探测器技术乳腺X线摄影的女性,共16006幅图像,用软件算法定量乳腺腺体含量并估算患者特异性平均腺体剂量(psMGD),并与厂家提供的器官剂量比较。结果显示校准后psMGD与器官剂量中位相对差异低于3%,但个体系统差异较大,器官剂量与psMGD偏差超过10%的图像比例在11%至46%之间。结论认为患者特异性平均腺体剂量(可能经系统校准调整)比厂家器官剂量能实现更准确的个体乳腺剂量学。
要点
- 患者特异性平均腺体剂量(psMGD)估算比厂家提供的器官剂量更准确。
- 系统校准调整对存在入射剂量明显校准问题的设备尤为有用。
- 诊断参考水平可基于psMGD或器官剂量制定。
诊断标准与关键数值
- 校准后psMGD与器官剂量的总体中位相对差异低于3%。
- 经校准psMGD与厂家器官剂量的平均差异为4.1%,范围-16.3%至+24.5%。
- 器官剂量与psMGD相比偏差超过10%的图像比例在11%至46%之间,取决于乳腺摄影系统。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To evaluate the impact of increasing levels of accuracy for mean glandular dose (MGD) evaluation in the era of the European Directive 2013/59/Euratom transposition.
METHOD: 4028 women who had a mammography examination by one of five mammography units using different detector technologies were included in this study. 16,006 images were processed by a software algorithm that determines breast glandularity quantitatively and uses this to estimate patient-specific MGD (psMGD). Entrance dose (ED) values and half value layers (HVLs) measured for each mammography system were collected to evaluate the effect of equipment calibration in psMGD calculation. The psMGD values adjusted for system calibration were compared with organ dose (OD) provided by manufacturers as image metadata.
RESULTS: Overall median relative difference between calibrated psMGD and organ dose was below 3%, with larger differences for individual systems. The psMGD adjustment for system calibration was particularly useful for one system for which ED had an evident miscalibration issue. The mean difference between psMGD with calibration and organ dose provided by manufacturers was 4.1 %, ranging from -16.3 % to +24.5 %. The proportion of images for which organ dose was more than 10 % 'inaccurate' compared to psMGD was between 11 % and 46 %, depending on the mammography system.
CONCLUSION: Patient-specific mean glandular dose, possibly adjusted for system calibration, allows more accurate individual breast dosimetry than what would be performed using organ dose provided by manufacturers. Conversely, definition of diagnostic reference levels could be achieved using either psMGD or organ dose.