文献速递技术与安全乳腺US

C-弹性成像:体外可行性体模研究

C-Elastography: In Vitro Feasibility Phantom Study.
PubMed 文献 · Ultrasound Med Biol · 2020-04-18

本文介绍一种新型超声技术C-弹性成像(CE),利用声辐射力产生的低频剪切波局部映射软组织非线性弹性,通过超声运动传感与三维黏弹性动力学模拟估算三阶弹性模量C,并在明胶体模内包埋的离体组织及仿组织包涵体上开展初步可行性实验,结果显示软、硬包涵体推点处平均C值分别为1.9和5.6,呈现明显而锐利的C对比。

要点

  • C-弹性成像利用150-250 Hz低频剪切波与三维黏弹性动力学模拟估算非线性模量C。
  • CE可在O(mm³)体积内测量C值,体模实验中软包涵体平均C为1.9、硬包涵体为5.6。
  • 该研究为体外可行性初步研究,尚需进一步验证其临床应用价值。

诊断标准与关键数值

  • C-弹性成像通过低频(150-250 Hz)剪切波局部映射软组织非线性弹性,基于声辐射力幅值与三阶弹性模量C呈线性关系的发现。
  • 软包涵体推点处平均C值为1.9,硬包涵体推点处平均C值为5.6。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

C-Elastography (CE) is a new ultrasound technique that locally maps the non-linear elasticity of soft tissue using low-frequency (150-250 Hz) shear waves generated by the acoustic radiation force (ARF). CE is based on a recent finding that the magnitude of the ARF in an isotropic tissue-like solid is related linearly to a third-order modulus of elasticity, C, which is responsible for the coupling between deviatoric and volumetric constitutive behaviors. The main objective of the work described here was to examine the feasibility of using and performance of C-elastography in differentiating and characterizing soft tissue via a pilot study on ex vivo tissue and tissue-mimicking inclusions cast in a gelatin block. In this vein, the CE technique deploys a combination of ultrasound motion sensing and 3-D visco-elastodynamic simulation to estimate the non-linear modulus C. As ultrasound focusing inherently confines the ARF to a small region, CE provides the means for measuring C within O(mm3) volumes. Equipped with such data analysis, we performed in vitro CE experiments on agar-based, xenograft and normal breast tissue samples embedded in a gelatin matrix. The compound C-elastograms indicate marked (and sharp) C-contrast, with average values of 1.9 and 5.6 at push points inside the featured soft and hard inclusions, respectively.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Shahraki DP, Kumar V, Ghavami S, et al. C-Elastography: In Vitro Feasibility Phantom Study. Ultrasound Med Biol. 2020. PMID: 32312548. doi:10.1016/j.ultrasmedbio.2020.02.005

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。