我们应忽略、随访还是活检?人工智能决策支持对乳腺超声病灶评估的影响
该多中心回顾性研究纳入900个乳腺超声病灶,由15名医师分别在仅用超声与超声联合人工智能决策支持两种模式下判读;联合决策支持时平均AUC由0.83升至0.87,阅片者间一致性提高,BI-RADS 3升至4A及以上的类别切换由13.6%降至10.8%(p=0.04)。
要点
- 超声联合人工智能决策支持时平均AUC为0.87,仅用超声时为0.83。
- 决策支持系统阳性似然比为1.98,高于除一名以外所有阅片者。
- 联合决策支持时BI-RADS 3升至4A及以上的类别切换占10.8%,仅用超声时占13.6%(p=0.04)。
关键影像表现/征象
- 人工智能系统(Koios DS for Breast,Koios Medical)评估图像并将病灶归为良性、可能良性、可疑、可能恶性四类之一。
- 每名阅片者判读两次:750例分别以仅用超声或超声联合决策支持的方式判读,4周后以相反方式复判。
- 另有150例在两次判读中以相同方式呈现。
诊断标准与关键数值
- 仅用超声时阅片者平均AUC为0.83(95%CI 0.78-0.89),超声联合决策支持时平均AUC为0.87(95%CI 0.84-0.90)。
- 决策支持系统的阳性似然比为1.98(95%CI 1.78-2.18)。
- 仅用超声时阅片者间Kendall τ-b为0.54(95%CI 0.53-0.55),超声联合决策支持时为0.68(95%CI 0.67-0.69)。
- 类别切换(定义为从BI-RADS 3类跨至BI-RADS 4A类及以上)在仅用超声时占13.6%,超声联合决策支持时占10.8%(p=0.04)。
临床应用与阅片要点
- 15名阅片者中有14名在超声联合决策支持时AUC高于仅用超声时。
- 决策支持系统的阳性似然比高于除一名以外所有阅片者的阳性似然比。
- 联合决策支持时阅片者内变异性改善。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE. The objective of this study was to assess the impact of artificial intelligence (AI)-based decision support (DS) on breast ultrasound (US) lesion assessment. MATERIALS AND METHODS. A multicenter retrospective review of 900 breast lesions (470/900 [52.2%] benign; 430/900 [47.8%] malignant) on US by 15 physicians (11 radiologists, two surgeons, two obstetrician/gynecologists). An AI system (Koios DS for Breast, Koios Medical) evaluated images and assigned them to one of four categories: benign, probably benign, suspicious, and probably malignant. Each reader reviewed cases twice: 750 cases with US only or with US plus DS; 4 weeks later, cases were reviewed in the opposite format. One hundred fifty additional cases were presented identically in each session. DS and reader sensitivity, specificity, and positive likelihood ratios (PLRs) were calculated as well as reader AUCs with and without DS. The Kendall τ-b correlation coefficient was used to assess intraand interreader variability. RESULTS. Mean reader AUC for cases reviewed with US only was 0.83 (95% CI, 0.78-0.89); for cases reviewed with US plus DS, mean AUC was 0.87 (95% CI, 0.84-0.90). PLR for the DS system was 1.98 (95% CI, 1.78-2.18) and was higher than the PLR for all readers but one. Fourteen readers had better AUC with US plus DS than with US only. Mean Kendall τ-b for US-only interreader variability was 0.54 (95% CI, 0.53-0.55); for US plus DS, it was 0.68 (95% CI, 0.67-0.69). Intrareader variability improved with DS; class switching (defined as crossing from BI-RADS category 3 to BI-RADS category 4A or above) occurred in 13.6% of cases with US only versus 10.8% of cases with US plus DS (p = 0.04). CONCLUSION. AI-based DS improves accuracy of sonographic breast lesion assessment while reducing inter- and intraobserver variability.
引用
- Mango VL, Sun M, Wynn RT, et al. Should We Ignore, Follow, or Biopsy? Impact of Artificial Intelligence Decision Support on Breast Ultrasound Lesion Assessment. AJR Am J Roentgenol. 2020. PMID: 32319794. doi:10.2214/AJR.19.21872
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。