1.5-T与3.0-T胎儿MRI的比吸收率与比能量剂量比较

Specific Absorption Rate and Specific Energy Dose: Comparison of 1.5-T versus 3.0-T Fetal MRI.
PubMed 文献 · Radiology · 2020-04-07

该回顾性研究比较了单中心2012年7月至2016年10月间1.5-T与3.0-T胎儿MRI的比吸收率(SAR)和比能量剂量(SED),共纳入3247次胎儿MRI扫描、93764个序列。结果显示多数序列的SAR、SED及累积SED在两种场强间等效,但3.0-T的二维T1加权 spoiled gradient-echo 和三维稳态自由进动序列可能需要修改以尽可能降低能量沉积。

要点

  • 1.5-T与3.0-T胎儿MRI在多数序列的SAR、SED及累积SED上等效。
  • 3.0-T的二维T1加权SPGR和三维SSFP序列可能需要修改以降低能量沉积。
  • 本研究为单中心回顾性研究,数据来自DICOM头文件。

核心结论

  • 胎儿1.5-T与3.0-T MRI检查在多数情况下具有等效的能量指标;3.0-T序列如二维T1加权SPGR和三维SSFP可能需要修改以尽可能降低传递给患者的能量。

研究设计与样本

  • 单中心回顾性研究,纳入2012年7月至2016年10月间所有1.5-T和3.0-T胎儿MRI检查,共2952名孕妇、3247次胎儿MRI扫描(1.5-T 2784次,86%;3.0-T 463次,14%),共93764个序列(1.5-T 81535个,87%;3.0-T 12229个,13%);扫描序列包括二维和三维稳态自由进动(SSFP)、单次激发快速自旋回波、二维和三维T1加权 spoiled gradient-echo(SPGR)及回波平面成像;从DICOM头文件获取SAR、SED、累积SED和采集时间;采用等效界值为0.5(Cohen d效应量)的双单侧检验,P < .05认为统计学等效。

诊断标准与关键数值

  • 1.5-T vs 3.0-T MRI序列:平均SAR 1.09 W/kg ± 0.69 vs 1.14 W/kg ± 0.61,平均SED 33 J/kg ± 27 vs 38 J/kg ± 26,平均累积SED 965 J/kg ± 408 vs 996 J/kg ± 366(P < .001),上述指标均等效。

临床应用与阅片要点

  • 作者指出,3.0-T序列如二维T1加权SPGR和三维SSFP可能需要修改以尽可能降低传递给患者的能量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Background MRI performed at 3.0 T offers greater signal-to-noise ratio and better spatial resolution than does MRI performed at 1.5 T; however, for fetal MRI, there are concerns about the potential for greater radiofrequency energy administered to the fetus at 3.0-T MRI. Purpose To compare the specific absorption rate (SAR) and specific energy dose (SED) of fetal MRI at 1.5 and 3.0 T. Materials and Methods In this retrospective study, all fetal MRI examinations performed with 1.5- and 3.0-T scanners at one institution between July 2012 and October 2016 were evaluated. Two-dimensional (2D) and three-dimensional (3D) steady-state free precession (SSFP), single-shot fast spin-echo, 2D and 3D T1-weighted spoiled gradient-echo (SPGR), and echo-planar imaging sequences were performed. SAR, SED, accumulated SED, and acquisition time were retrieved from the Digital Imaging and Communications in Medicine header. Data are presented as mean ± standard deviation. Two one-sided tests with equivalence bounds of 0.5 (Cohen d effect size) were performed, with statistical equivalence considered at P < .05. Results A total of 2952 pregnant women were evaluated. Mean maternal age was 30 years ± 6 (age range, 12-49 years), mean gestational age was 24 weeks ± 6 (range, 17-40 weeks). A total of 3247 fetal MRI scans were included, with 2784 (86%) obtained at 1.5 T and 463 (14%) obtained at 3.0 T. In total, 93 764 sequences were performed, with 81 535 (87%) performed at 1.5 T and 12 229 (13%) performed at 3.0 T. When comparing 1.5- with 3.0-T MRI sequences, mean SAR (1.09 W/kg ± 0.69 vs 1.14 W/kg ± 0.61), mean SED (33 J/kg ± 27 vs 38 J/kg ± 26), and mean accumulated SED (965 J/kg ± 408 vs 996 J/kg ± 366, P < .001) were equivalent. Conclusion Fetal 1.5- and 3.0-T MRI examinations were found to have equivalent energy metrics in most cases. The 3.0-T sequences, such as two-dimensional T1-weighted spoiled gradient-echo and three-dimensional steady-state free precession, may require modification to keep the energy delivered to the patient as low as possible. © RSNA, 2020 Online supplemental material is available for this article.