两种迭代重建算法MBIR与ASIR在儿童低辐射剂量和低对比剂剂量腹部CT中的性能评估
本研究纳入59例接受低辐射剂量(100kVp)和低对比剂剂量(270 mgI/ml)腹部增强CT的儿童,比较MBIR、不同比例ASIR与FBP重建的图像质量。MBIR与ASIR100%的总体噪声和结构显示能力评分满足诊断要求,其余重建未达诊断要求;MBIR较FBP噪声降低62.86%–65.73%、CNR提高151.38%–170.69%。结论认为MBIR或ASIR100%可改善图像质量以满足诊断要求,MBIR表现最佳。
要点
- MBIR的总体图像噪声评分和结构显示能力评分均为4.00±0.00,满足临床诊断要求。
- ASIR100%的总体噪声评分为3.27±0.33、结构显示能力评分为3.25±0.43,满足诊断要求。
- MBIR较FBP噪声降低62.86%–65.73%,CNR提高151.38%–170.69%(P<0.05)。
核心结论
- MBIR或ASIR100%可改善儿童低辐射剂量和低对比剂剂量腹部CT的图像质量以满足诊断要求,其中MBIR表现最佳。
研究设计与样本
- 本研究为影像重建算法性能评估研究,纳入59例儿童(男32例、女27例),中位年龄4.0岁(范围0.3–13岁)。
- 所有患儿接受低辐射剂量(100kVp)和低对比剂剂量(270 mgI/ml)腹部增强CT。
- 原始数据分别用MBIR、ASIR和FBP重建为6组:MBIR、100%ASIR、80%ASIR、60%ASIR、40%ASIR和FBP。
- 测量肝、胰腺、肾和腹主动脉的CT值、标准差、信噪比(SNR)和对比噪声比(CNR),并由两名放射科医师独立以4分制主观评价总体图像噪声和结构显示能力(3分为临床可接受)。
- 重建组间测量值比较采用单因素方差分析。
诊断标准与关键数值
- 主观图像质量采用4分制评价总体图像噪声和结构显示能力,3分为临床可接受。
- MBIR的总体图像噪声评分为4.00±0.00、结构显示能力评分为4.00±0.00,满足诊断要求。
- ASIR100%的总体图像噪声评分为3.27±0.33、结构显示能力评分为3.25±0.43,满足诊断要求;其余重建未达诊断要求。
- 与FBP图像相比,MBIR图像各器官噪声降低62.86%–65.73%(F=48.15–80.47,P<0.05),CNR提高151.38%–170.69%(F=22.94–38.02,P<0.05)。
临床应用与阅片要点
- 在儿童低辐射剂量(100kVp)和低对比剂剂量(270 mgI/ml)腹部增强CT中,MBIR或ASIR100%重建的图像可满足诊断要求,MBIR表现最佳。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: The adverse effect of low-dose CT on image quality may be mitigated using iterative reconstructions. The purpose of this study was to evaluate the performance of the full model-based iterative reconstruction (MBIR) and adaptive statistical reconstruction (ASIR) algorithms in low radiation dose and low contrast dose abdominal contrast-enhanced CT (CECT) in children.
METHODS: A total of 59 children (32 males and 27 females) undergoing low radiation dose (100kVp) and low contrast dose (270 mgI/ml) abdominal CECT were enrolled. The median age was 4.0 years (ranging from 0.3 to 13 years). The raw data were reconstructed with MBIR, ASIR and filtered back-projection (FBP) algorithms into 6 groups (MBIR, 100%ASIR, 80%ASIR, 60%ASIR, 40%ASIR and FBP). The CT numbers, standard deviations, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of liver, pancreas, kidney and abdominal aorta were measured. Two radiologists independently evaluated the subjective image quality including the overall image noise and structure display ability on a 4-point scale with 3 being clinically acceptable. The measurements among the reconstruction groups were compared using one-way ANOVA.
RESULTS: The overall image noise score and display ability were 4.00 ± 0.00 and 4.00 ± 0.00 with MBIR and 3.27 ± 0.33 and 3.25 ± 0.43 with ASIR100%, respectively, which met the diagnostic requirement; other reconstructions couldn't meet the diagnostic requirements. Compared with FBP images, the noise of MBIR images was reduced by 62.86-65.73% for the respective organs (F = 48.15-80.47, P < 0.05), and CNR increased by 151.38-170.69% (F = 22.94-38.02, P < 0.05).
CONCLUSIONS: MBIR or ASIR100% improves the image quality of low radiation dose and contrast dose abdominal CT in children to meet the diagnostic requirements, and MBIR has the best performance.
引用
- Sun J, Yang L, Zhou Z, et al. Performance evaluation of two iterative reconstruction algorithms, MBIR and ASIR, in low radiation dose and low contrast dose abdominal CT in children. Radiol Med. 2020. PMID: 32319005. doi:10.1007/s11547-020-01191-1
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。