斯洛文尼亚放疗CT模拟定位程序国家诊断参考水平的建立
该研究基于全国最大放疗科两台CT单元4个月内1631例患者数据,计算腹部、乳腺、胸部、头部、头颈部、盆腔和脊柱模拟定位的总DLP和CTDIvol,以第三四分位数确定国家诊断参考水平。结果显示两台CT单元间剂量值存在显著差异,部分协议暴露参数应审查优化。
要点
- 放疗CT模拟定位的国家DRL以第三四分位数DLP和CTDIvol表示,覆盖7类解剖部位。
- 头部模拟DLP第三四分位数为1942.4 mGy∙cm、CTDIvol为76.9 mGy,均为所列部位中最高。
- 两台CT单元间剂量值存在显著差异,提示需审查并优化部分协议的暴露参数。
诊断标准与关键数值
- 腹部、乳腺、胸部、头部、头颈部、盆腔和脊柱的DLP第三四分位数分别为1116.2、606.6、832.4、1942.4、969.2、677.1和1042.4 mGy∙cm。
- 同一序列程序的CTDIvol第三四分位数分别为18.7、13.3、19.2、76.9、22.6、17.9和22.2 mGy。
- 国家DRL以两台CT单元合并数据的第三四分位数确定。
临床应用与阅片要点
- 放疗CT模拟定位的剂量参考水平可按解剖部位分别报告DLP和CTDIvol第三四分位数。
- 同一机构内不同CT单元间剂量值平均存在显著差异,需审查并优化暴露参数。
局限性
- 原文结论指出两台CT单元在某些协议间存在差异,暴露参数应审查和优化。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To propose national diagnostic reference levels (DRLs) for radiotherapy (RT) computed tomography (CT) localization purposes, compare both CT units used in the largest RT department in the country and to compare gathered results with other published DRLs in order to discover any need of optimization.
METHODS: In total, 1631 patient data (time spend of 4 months) regarding sex, examination type, total dose-length product (DLP) and CTDIvol was collated on two CT units. Those simulation procedures account for more than 80 % of all simulation procedures performed nationwide. Then, total DLP and CTDIvol was calculated and mean, median and 3rd quartile for both units together were presented to determine national DRLs for simulation procedures. The same data was later compared between both units to discover any potential need for optimization.
RESULTS: 3rd quartile values of DLP for abdomen, breast, chest, head, head and neck, pelvis and spine were 1116.2, 606.6, 832.4, 1942.4, 969.2, 677.1 and 1042.4 mGy∙cm, respectively. 3rd quartile CTDIvol values for the same sequence of procedures were 18.7, 13.3, 19.2, 76.9, 22.6, 17.9 and 22.2 mGy, respectively. Among the two units, the mentioned dose values were on average significantly higher on one CT unit than on the other unit.
CONCLUSIONS: When comparing collected dose values with other studies, RT CT DRLs showed that radiation doses from our institution were similar or even lower. Some variations were found between both CT units in certain protocols, so exposure parameters should be reviewed and optimized.
引用
- Zalokar N, Žager Marciuš V, Mekiš N, et al. Establishment of national diagnostic reference levels for radiotherapy computed tomography simulation procedures in Slovenia. Eur J Radiol. 2020. PMID: 32302897. doi:10.1016/j.ejrad.2020.108979
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。