弹簧圈栓塞颅内动脉瘤远端疑似金属栓塞可被磁敏感加权MR成像检出
回顾性分析本机构2018年9月至2019年3月700例弹簧圈栓塞颅内动脉瘤术后随访MR图像,用SWI评估远端圆形低信号病灶的发生频率、可能原因及临床意义。20例(2.8%)共25个病灶,直径5–11 mm(中位8 mm),除2个外均位于动脉瘤下游供血区;受影响患者均未出现局灶性神经症状,作者认为此类病灶似乎无临床影响。
要点
- 弹簧圈栓塞颅内动脉瘤术后SWI随访可见远端圆形低信号病灶,本组发生率为2.8%(20/700例),共25个病灶。
- 病灶直径5–11 mm(中位8 mm),除2个外均位于动脉瘤下游血管供血区,且均位于置入导引导管的血管供血区。
- 作者认为此类疑似金属栓塞病灶似乎无临床影响,受累患者均未出现局灶性神经症状。
核心结论
- 弹簧圈栓塞颅内动脉瘤远端金属颗粒栓塞在随访MR成像中偶尔可见,起源尚不清楚。
- 此类SWI可检出的病灶似乎无临床影响,受累患者均未出现局灶性神经症状。
研究设计与样本
- 回顾性分析本机构2018年9月至2019年3月间700例弹簧圈栓塞颅内动脉瘤术后随访MR成像研究。
- 在出现病灶的患者中筛查手术记录中使用的器械,包括弹簧圈、微导管、微导丝和支架,并查阅病历判断是否出现相关症状。
关键影像表现/征象
- SWI上表现为圆形低信号病灶,直径5–11 mm(中位8 mm),>5 mm的圆形低信号病灶被推定为金属性。
- 25个病灶中除2个外均位于动脉瘤下游脑区,但全部位于置入导引导管的血管供血区。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: After endovascular coiling of intracranial aneurysms, round dark parenchymal lesions believed to be particulate metal are sometimes encountered in MR imaging studies of the brain. We used SWI to assess the frequency of such occurrences, in addition to exploring likely causes and clinical implications.
MATERIALS AND METHODS: We reviewed 700 MR imaging studies performed between September 2018 and March 2019 at our institution as follow-up monitoring of coiled intracranial aneurysms. Any sizeable (>5 mm) rounded dark-signal lesions encountered were presumed to be metallic. The magnitudes and locations of such lesions were recorded. In patients with these lesions, pertinent procedural documentation was screened for devices used, including coils, microcatheters, microguidewires, and stents. Medical records were also examined to determine whether any related symptoms ensued.
RESULTS: Twenty patients (2.8%) exhibited a total of 25 lesions on SWI. Diameters ranged from 5 to 11 mm (median, 8 mm). All except 2 lesions were located in brain regions downstream from aneurysms, but all lesions occupied vascular territories of vessels used to place guiding catheters. Other than the Synchro 14, which was routinely deployed, no device was regularly used in patients with SWI-detectable lesions; and none of the affected patients developed focal neurologic symptoms as a consequence.
CONCLUSIONS: Although the origins remain unclear, distal embolization of particulate metal distal to coiled cerebral aneurysms is occasionally observed on follow-up MR imaging studies. Such lesions, however, seem to have no apparent clinical impact.
引用
- Yoo DH, Cho YD, Lee HS, et al. Suspected Metallic Embolization Distal to Coiled Intracranial Aneurysms Detectable by Susceptibility-Weighted MR Imaging. AJNR Am J Neuroradiol. 2020. PMID: 32273325. doi:10.3174/ajnr.A6506
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。