文献速递心血管介入技术与安全DSA

利用眼动追踪技术量化介入透视中可避免的辐射暴露

Quantification of Avoidable Radiation Exposure in Interventional Fluoroscopy With Eye Tracking Technology.
PubMed 文献 · Invest Radiol · 2020-01-01

研究用移动眼动追踪眼镜测量术者在心血管介入中注视透视屏的时间,并以机器学习算法自动计算可避免辐射比例。30例介入由5名术者完成,平均43.5%的辐射可避免,相当于每例平均229 mGy空气比释动能和32,781 mGycm剂量面积乘积,或超过11分钟的可避免X线使用。

要点

  • 术者未注视透视屏时X线仍开启,平均43.5%的辐射暴露可避免。
  • 每例平均可避免229 mGy空气比释动能和32,781 mGycm剂量面积乘积。
  • 每例平均可避免超过11分钟的X线使用时间。

核心结论

  • 心血管介入中相当比例的辐射暴露实际上可避免(平均43.5%,t29=18.86,P<0.00001)。

研究设计与样本

  • 研究纳入30例心血管介入,由5名不同术者完成,使用移动眼动追踪眼镜测量术者注视透视屏的时机。
  • 采用新型机器学习算法和图像处理技术自动分析数据并计算可避免辐射百分比。
  • 基于该百分比计算潜在可避免的辐射剂量。

诊断标准与关键数值

  • 可避免辐射定义为术者在X线开启时未注视透视屏,平均(标准差)为43.5%(12.6%)。
  • 可避免空气比释动能平均(标准差)为229(66)mGy,可避免剂量面积乘积平均(标准差)为32,781(9420)mGycm。
  • 每例可避免X线使用时间超过11分钟。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Reducing avoidable radiation exposure during medical procedures is a top priority. The purpose of this study was to quantify, for the first time, the percentage of avoidable radiation during fluoroscopically guided cardiovascular interventions using eye tracking technologies.

MATERIALS AND METHODS: Mobile eye tracking glasses were used to measure precisely when the operators looked at a fluoroscopy screen during the interventions. A novel machine learning algorithm and image processing techniques were used to automatically analyze the data and compute the percentage of avoidable radiation. Based on this percentage, the amount of potentially avoidable radiation dose was computed.

RESULTS: This study included 30 cardiovascular interventions performed by 5 different operators. A significant percentage of the administered radiation (mean [SD], 43.5% [12.6%]) was avoidable (t29 = 18.86, P < 0.00001); that is, the operators were not looking at the fluoroscopy screen while the x-ray was on. On average, this corresponded to avoidable amounts of air kerma (mean [SD], 229 [66] mGy) and dose area product (mean [SD], 32,781 [9420] mGycm), or more than 11 minutes of avoidable x-ray usage, per procedure.

CONCLUSIONS: A significant amount of the administered radiation during cardiovascular interventions is in fact avoidable.

来源:Invest Radiol,查看原文。版权归原作者所有。