双能对比增强数字乳腺摄影与数字乳腺断层摄影的平均腺体剂量评估及其与压迫乳腺厚度的关系研究

Evaluation of average glandular dose and investigation of the relationship with compressed breast thickness in dual energy contrast enhanced digital mammography and digital breast tomosynthesis.
PubMed 文献 · Eur J Radiol · 2020-03-03

研究记录135例采用自动曝光控制模式的乳腺摄影检查,计算双能对比增强数字乳腺摄影(CEDM)与数字乳腺断层摄影(DBT)的平均腺体剂量(AGD),并分析AGD与压迫乳腺厚度(CBT)、压迫力(CF)的关系。结果显示两种技术在CC和MLO视图的剂量均符合乳腺摄影最大剂量推荐,CEDM的AGD中位数在各视图均显著低于DBT,AGD与CBT呈正相关和线性回归,而CF与AGD无相关性。

要点

  • CEDM的AGD中位数在CC和MLO视图均显著低于DBT(p << 0.01)。
  • AGD与CBT呈正相关和线性回归,MLO视图R²在CEDM为0.74、DBT为0.61。
  • CF与AGD无相关性,两种技术剂量均符合乳腺摄影最大剂量推荐。

诊断标准与关键数值

  • CEDM的AGD中位数在CC和MLO视图均显著低于DBT(p << 0.01)。
  • AGD与CBT呈正相关和线性回归,MLO视图R²在CEDM为0.74、DBT为0.61。
  • CF与AGD无相关性及线性回归。
  • Kruskal Wallis检验显示CEDM与DBT的AGD在CC视图与MLO视图之间差异有统计学意义(p < 0.01)。
  • CBT范围为23至94 mm,CC视图厚度中位值52 mm、MLO视图57 mm。

临床应用与阅片要点

  • 两种技术剂量均符合乳腺摄影最大剂量推荐,阅片时可关注不同视图的AGD差异。
  • CBT可影响AGD,CF与AGD不相关,报告剂量相关质控信息时可参考CBT。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To quantitatively assess the dose of Dual energy contrast enhanced digital mammography (CEDM) and digital breast tomosynthesis (DBT) and to investigate the relationship between average absorbed glandular dose (AGD), compressed breast thickness (CBT) and compression force (CF).

MATERIALS AND METHODS: All CEDM and DBT examinations were performed in cranio-caudal (CC) and medio-lateral oblique (MLO) view. Exposure parameters of 135 mammographic procedures that using AEC (automatic exposure control) mode were recorded. AGDs were calculated. Kruskal Wallis test was performed.

RESULTS: CBT population ranged from 23 to 94 mm with a thickness median value of 52 mm in CC view and of 57 mm in MLO views. CEDM AGD median value was significatively lower than DBT AGD in each views (p << 0.01). AGD showed a positive correlation and linear regression with CBT for both CEDM and DBT while CF did not show a correlation and linear regression with AGD. The highest values were found for MLO view: R2 of 0.74 for CEDM and R2 of 0.61 for DBT. Kruskal Wallis test shows that there was a difference statistically significant between AGD values of CEDM and DBT in CC view respect to MLO views (p < 0.01).

CONCLUSIONS: Dose values of both techniques meet the recommendations for maximum dose in mammography. The results of the present study indicated that there was significant difference between AGD for CEDM and DBT exposure in different views (AGD in CC views had the lowest value) and that CBT could influence the AGD while CF was not correlated to AGD.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Fusco R, Raiano N, Raiano C, et al. Evaluation of average glandular dose and investigation of the relationship with compressed breast thickness in dual energy contrast enhanced digital mammography and digital breast tomosynthesis. Eur J Radiol. 2020. PMID: 32151787. doi:10.1016/j.ejrad.2020.108912