对比增强超声对不同大小乳腺病变的诊断效能:与磁共振成像的比较研究
本研究纳入406例经常规超声检出乳腺实性肿块的406例患者,均行对比增强超声(CEUS)和MRI检查,以组织学为参考标准,按病灶大小分组比较诊断效能。CEUS在各组中检出乳腺癌的敏感度与MRI相似,但MRI的特异度、准确率和ROC曲线下面积均优于CEUS;两种方法联合可显著提高CEUS单独诊断的效能,但除>20 mm组外,联合诊断的特异度和准确率并不优于MRI单独诊断。
要点
- CEUS检出乳腺癌的敏感度与MRI相似,三组比较P值均为1.0000。
- MRI的特异度、准确率和ROC曲线下面积(Az)均优于CEUS,所有比较P<0.05。
- 联合诊断可显著提高CEUS单独诊断的效能(P<0.05),但除>20 mm组外特异度和准确率不优于MRI单独诊断。
核心结论
- CEUS检出乳腺癌具有良好的敏感度,与MRI联合使用可优化乳腺癌预测的诊断特异度和准确率。
- 除>20 mm组外,联合诊断的特异度和准确率并不优于MRI单独诊断。
研究设计与样本
- 本研究为比较性研究,纳入406例患者的406个经常规超声检出的乳腺实性肿块,均行CEUS和MRI检查,以组织学结果为参考标准。
- 病灶按大小分为三组:Group 1为≤20 mm,Group 2为>20 mm,Group 3为全部病灶。
诊断标准与关键数值
- 采用敏感度、特异度、准确率和受试者工作特征(ROC)曲线分析评估各影像方法的诊断效能。
- 组织学诊断显示良性病灶194个、恶性病灶212个。
- CEUS检出乳腺癌的敏感度与MRI相似,三组比较P值均为1.0000。
- MRI的特异度、准确率和ROC曲线下面积(Az)值均优于CEUS,所有比较P<0.05。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: This study aimed to compare the diagnostic performance of contrast-enhanced ultrasound (CEUS), MRI, and the combined use of the two modalities for differentiating breast lesions of different sizes.
METHODS: A total of 406 patients with 406 solid breast masses detected by conventional ultrasound underwent both CEUS and MRI scans. Histological results were used as reference standards. The lesions were categorized into three groups according to size (Group 1, ≤ 20 mm; Group 2, > 20 mm, Group 3: total lesions). Sensitivity, specificity, accuracy, and receiver operating characteristic (ROC) curve analysis were used to assess the diagnostic performance of these imaging methods for breast lesions.
RESULTS: There were 194 benign and 212 malignant breast lesions according to the histological diagnosis. Compared with MRI, CEUS demonstrated similar sensitivity in detecting breast cancer (p = 1.0000 for all) in all the three groups. With regard to specificity, accuracy, and the area under the ROC curve (Az) values, MRI showed a better performance than that shown by CEUS (p <0.05 for all), and the combination of the two modalities improved the diagnostic performance of CEUS alone significantly (p <0.05 for all) in all the three groups. However, the diagnostic specificity and accuracy of the combined method was not superior to that of MRI alone except for Group 2.
CONCLUSION: CEUS demonstrated good sensitivity in detecting breast cancer, and the combined use with MRI can optimize the diagnostic specificity and accuracy in breast cancer prediction.
ADVANCES IN KNOWLEDGE: Few studies have compared the diagnostic efficacy of CEUS and MRI, and this study is the first attempt to seek out the diagnostic values for breast lesions of variable sizes (lesions with ≤20 mm and >20 mm).