光子计数探测器螺旋乳腺CT的病灶可检出性与辐射剂量:体模研究
该体模研究评估了配备光子计数探测器的临床螺旋乳腺CT在不同管电流(5–125 mA)和两种板厚(100与160 mm)下的病灶可检出性、图像质量与平均吸收剂量。结果显示,290与400 μm微钙化及3.18–6.32 mm肿块在所有管电流下均可见;1.8 mm软组织肿块在25 mA及以上可见;196 μm微钙化在小板厚25 mA及以上、大板厚40 mA及以上可检出。小乳腺与大乳腺在25与40 mA时平均剂量分别为4.30±0.01与5.70±0.02 mGy。
要点
- 290与400 μm微钙化及3.18、4.76、6.32 mm肿块在5–125 mA所有管电流下均可见。
- 1.8 mm软组织肿块在25 mA及以上管电流下可见。
- 196 μm微钙化在小板厚25 mA及以上、大板厚40 mA及以上可检出。
诊断标准与关键数值
- 290与400 μm微钙化及3.18、4.76、6.32 mm肿块在5–125 mA所有管电流下均可见。
- 1.8 mm软组织肿块在25 mA及以上管电流下可见。
- 196 μm微钙化在小板厚于25 mA及以上、大板厚于40 mA及以上可检出。
- 小板厚乳腺在25 mA时平均吸收剂量为4.30±0.01 mGy,大板厚乳腺在40 mA时平均吸收剂量为5.70±0.02 mGy。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The aim of the article was to evaluate the lesion detectability, image quality, and radiation dose of a dedicated clinical spiral breast computed tomography (CT) system equipped with a photon-counting detector, and to propose optimal scan parameter settings to achieve low patient dose levels and optimal image quality.
METHODS: A breast phantom containing inserts mimicking microcalcifications (diameters 196, 290, and 400 μm) and masses (diameters 1.8, 3.18, 4.76, and 6.32 mm) was examined in a spiral breast CT system with systematic variations of x-ray tube currents between 5 and 125 mA, using 2 slabs of 100 and 160 mm. Signal-to-noise ratio and contrast-to-noise ratio measurements were performed by region of interest analysis. Two experienced radiologists assessed the detectability of the inserts. The average absorbed dose was calculated in Monte Carlo simulations.
RESULTS: Microcalcifications in diameters of 290 and 400 μm and masses in diameters of 3.18, 4.76, and 6.32 mm were visible for all tube currents between 5 and 125 mA. Soft tissue masses in a diameter of 1.8 mm were visible at tube currents of 25 mA and higher. Microcalcifications with a diameter of 196 μm were detectable at a tube current of 25 mA and higher in the small, and at a tube current of 40 mA and higher in the large slab. For the small and large breast, at a tube current of 25 and 40 mA, an average dose value of 4.30 ± 0.01 and 5.70 ± 0.02 mGy was calculated, respectively.
CONCLUSIONS: Optimizing tube current of spiral breast CT according to the breast size enables the visualization of microcalcifications as small as 196 μm while keeping dose values in the range of conventional mammography.