利用第11胸椎最小化CT-KUB扫描野
该审计研究评估了输尿管结石CT-KUB扫描的过度扫描长度,并采用以第11胸椎为上界的新方案进行整改。第一阶段90例扫描平均过度扫描百分比为28.2%,第二阶段105例降至10.6%,且第二阶段100%扫描完整包含双肾。
要点
- 以第11胸椎作为CT-KUB扫描上界可有效减少过度扫描。
- 整改后平均过度扫描百分比从28.2%降至10.6%,且全部扫描完整包含双肾。
- 过度扫描源于扫描技术不一致,标准化流程可最小化患者剂量而不影响图像质量。
核心结论
- 过度扫描源于CT-KUB扫描技术不一致;通过可重复的方法标准化流程后,过度扫描目标轻松达成,患者剂量最小化且未影响扫描质量。
研究设计与样本
- 回顾性审计研究,第一阶段连续纳入90例因输尿管结石行CT-KUB的扫描,第二阶段(3个月后)纳入105例;使用IMPAX软件分析。
诊断标准与关键数值
- 第一阶段过度扫描>15%的扫描占94.4%,平均过度扫描百分比为28.2%,59%的扫描上界在T10或以下,99%的扫描完整包含双肾;第二阶段平均过度扫描百分比降至10.6%(标准差=4.4%),过度扫描影响35.2%的扫描,8%的扫描上界在T10或以下,100%的扫描完整包含双肾。
临床应用与阅片要点
- 作者采用以第11胸椎作为扫描上界的新方案并传达给放射技师,该审计显示该标准化方案可行,可用于最小化患者过度扫描。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Computed tomography scans of the kidney, ureters, and bladder (CT-KUB) are crucial in investigating urinary calculi but impart a substantial radiation doses. Radiation can be limited by minimising the scanning field to the necessary area (i.e. from the kidneys to urethra). Before auditing, the superior limit of CT-KUB scans had not been formally clarified at our trust. Consistently ensuring the upper limit of scans is at or below T10 has been shown to be a viable method of performing CT-KUB scans. This study aimed to assess the overscan length of CT-KUB investigations and modify practice accordingly to minimise it. There were two standards that were set for CT-KUB scanning. First, the mean percentage overscan length (i.e. percentage of the scan above the kidneys) should be <15%. Second, all scans should include the superior borders of both kidneys.
METHODS: 90 consecutive CT-KUB scans for ureteric calculus were retrospectively investigated using IMPAX software in the first phase of data collection. After these data were analysed, a newly devised protocol using T11 as the superior scan limit was delivered to radiographers in the department. and 105 in the second phase (re-audit). The analysis parameters were: percentage overscan length, distance between diaphragm and upper border of kidneys, vertebral level at which the scan commenced, and whether both kidneys were fully included.
RESULTS: In the first phase, overscan of >15% was present in 94.4% of scans. The mean percentage overscan length was 28.2%. The superior vertebral limit of 59% of scans was at T10 or below and a lower superior vertebral limit correlated with decreasing overscan. 99% of scans fully included both kidneys. In the second phase (3 months later), the mean overscan percentage reduced to 10.6% (standard deviation = 4.4%). Excessive overscan affected 35.2% of scans. The superior vertebral limit of 8% of scans was at T10 or below. 100% of scans fully included both kidneys.
CONCLUSION: Excessive overscanning was due to inconsistent technique in capturing CT-KUB scans. Before this audit, the superior limit of CT-KUB scans had not been formally clarified at our trust. By successfully standardising the process with a reproducible method, the overscan target was comfortably met. Therefore, patient dose was minimised without compromising scan quality.
ADVANCES IN KNOWLEDGE: This audit has successfully shown a feasible standardised protocol for CT-KUB investigations which can be used to minimise overscanning of patients.