文献速递泌尿生殖腹部MR

前列腺MRI质量:PI-RADS标准是否足够?

Quality of Prostate MRI: Is the PI-RADS Standard Sufficient?
PubMed 文献 · Acad Radiol · 2020-03-04

研究纳入62例来自62家机构的前列腺MRI,由6名读者评估图像充分性和质量,并依据DICOM元数据评估对PI-RADSv2最低技术标准的依从性。结果显示依从性与图像质量仅T2W有弱相关,与DWI质量无统计学关系,依从PI-RADSv2最低技术标准并未提高图像质量充分性的可能性。

要点

  • 多数前列腺MRI图像被评為充分,但T2W和DWI的充分率分别为83%和61%,DWI充分率明显较低。
  • 对PI-RADSv2最低技术标准的依从性与T2W图像质量仅呈弱相关(tau-b=0.22,p=0.01)。
  • 遵循全部7条T2W规则的图像质量中位评分更高(3.58 vs 3.0,p=0.012)。

核心结论

  • 在62家开展前列腺MRI的机构中,少数图像被认为高质量,多数被认为充分;DWI的充分研究比例明显较低;依从PI-RADSv2最低技术标准并未增加获得质量充分T2W或DWI的可能性。

研究设计与样本

  • 研究纳入2017年9月至2018年9月12个月期间连续转诊至本中心的62例前列腺MRI检查(含T2W和DWI),来自62家不同机构;6名读者评估图像充分性并采用1-5分进行图像质量评分;将PI-RADSv2最低技术标准分别归纳为T2W的7条规则和DWI的10条规则;使用DICOM元数据评估依从性;采用Kendall Rank tau-b分析读者质量评分,采用Wilcoxon检验分析读者充分性评分与依从性的关联。

关键影像表现/征象

  • 62例图像中,52例(83%)T2W和38例(61%)DWI被多数读者评为充分。
  • 读者充分性评分与对PI-RADSv2最低技术标准的依从性无显著关联。
  • T2W图像质量与依从性呈弱相关(tau-b=0.22,p=0.01)。
  • 遵循全部7条PI-RADSv2 T2W规则的图像质量中位平均评分更高(7/7组3.58 vs <7/7组3.0,p=0.012)。
  • DWI质量与对PI-RADSv2最低技术标准的依从性无统计学关系。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVE: The Prostate Imaging Reporting and Data System version 2 (PI-RADSv2) published a set of minimum technical standards (MTS) to improve image quality and reduce variability in multiparametric prostate MRI. The effect of PIRADSv2 MTS on image quality has not been validated. We aimed to determine whether adherence to PI-RADSv2 MTS improves study adequacy and perceived quality.

MATERIALS AND METHODS: Sixty-two prostate MRI examinations including T2 weighted (T2W) and diffusion weighted image (DWI) consecutively referred to our center from 62 different institutions within a 12-month period (September 2017 to September 2018) were included. Six readers assessed images as adequate or inadequate for use in PCa detection and a numerical image quality ranking was given using a 1-5 scale. The PI-RADSv2 MTS were synthesized into sets of seven and 10 rules for T2W and DWI, respectively. Image adherence was assessed using Digital Imaging and Communications in Medicine (DICOM) metadata. Statistical analysis of survey results and image adherence was performed based on reader quality scoring (Kendall Rank tau-b) and reader adequate scoring (Wilcoxon test for association) for T2 and DWI quality assessment.

RESULTS: Out of 62 images, 52 (83%) T2W and 38 (61%) DWIs were rated to be adequate by a majority of readers. Reader adequacy scores showed no significant association with adherence to PI-RADSv2. There was a weak (tau-b = 0.22) but significant (p value = 0.01) correlation between adherence to PIRADSv2 MTS and image quality for T2W. Studies following all PI-RADSv2 T2W rules achieved a higher median average quality score (3.58 for 7/7 vs. 3.0 for <7/7, p = 0.012). No statistical relationship with PI-RADSv2 MTS adherence and DWI quality was found.

CONCLUSION: Among 62 sites performing prostate MRI, few were considered of high quality, but the majority were considered adequate. DWI showed considerably lower rates of adequate studies in the sample. Adherence to PI-RADSv2 MTS did not increase the likelihood of having a qualitatively adequate T2W or DWI.

来源:Acad Radiol,查看原文。版权归原作者所有。