首过心肌灌注磁共振成像中因视觉错觉产生的暗环伪影感知

Perceived Dark Rim Artifact in First-Pass Myocardial Perfusion Magnetic Resonance Imaging Due to Visual Illusion.
PubMed 文献 · Korean J Radiol · 2020-01-01

研究通过数值模体与健康志愿者心脏灌注图像,证明在k空间窗技术去除定量信号下凹后,受试者仍感知到左心室-心肌边界暗带,提示视觉错觉是暗环伪影的另一潜在原因;彩色标度减少亮度变化后该错觉暗带不可见。

要点

  • k空间窗技术去除定量信号下凹后,暗环仍被感知,提示视觉错觉可致暗环伪影。
  • 暗环在左心室峰值强化时比心肌峰值强化时更明显,因左心室与心肌强度差更大。
  • 采用减少亮度变化的彩色标度显示时,数值模体和在体图像中的错觉暗带均不可见。

关键影像表现/征象

  • k空间窗处理后,模体图像中左心室-心肌边界附近信号强度呈单调递减,边界附近有限差分值为负,范围-1.07至-0.14。
  • 尽管定量信号下凹已被去除,暗带仍被感知,提示所感知的暗环为视觉错觉。
  • 在体图像经k空间窗处理去除定量信号下凹后仍感知暗环。
  • 左心室峰值强化时感知的暗环比心肌峰值强化时更强,因左心室与心肌间强度差更大。
  • 数值模体和在体图像中,减少亮度变化的彩色图显示时错觉暗带均不可见。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To demonstrate that human visual illusion can contribute to sub-endocardial dark rim artifact in contrast-enhanced myocardial perfusion magnetic resonance images.

MATERIALS AND METHODS: Numerical phantoms were generated to simulate the first-passage of contrast agent in the heart, and rendered in conventional gray scale as well as in color scale with reduced luminance variation. Cardiac perfusion images were acquired from two healthy volunteers, and were displayed by the same gray and color scales used in the numerical study. Before and after k-space windowing, the left ventricle (LV)-myocardium boarders were analyzed visually and quantitatively through intensity profiles perpendicular the boarders.

RESULTS: k-space windowing yielded monotonically decreasing signal intensity near the LV-myocardium boarder in the phantom images, as confirmed by negative finite difference values near the board ranging -1.07 to -0.14. However, the dark band still appears, which is perceived by visual illusion. Dark rim is perceived in the in-vivo images after k-space windowing that removed the quantitative signal dip, suggesting that the perceived dark rim is a visual illusion. The perceived dark rim is stronger at peak LV enhancement than the peak myocardial enhancement, due to the larger intensity difference between LV and myocardium. In both numerical phantom and in-vivo images, the illusory dark band is not visible in the color map due to reduced luminance variation.

CONCLUSION: Visual illusion is another potential cause of dark rim artifact in contrast-enhanced myocardial perfusion MRI as demonstrated by illusory rim perceived in the absence of quantitative intensity undershoot.