植入非MRI兼容心脏起搏器的绵羊颈椎和胸部MRI的临床与组织病理学结果

Clinical and histopathological outcome of cervical and chest MRI involving non-MRI-conditional cardiac pacemakers: a study using sheep models.
PubMed 文献 · Radiol Med · 2020-03-21

在两只雄性绵羊右心室心尖或右心房植入两种双腔非MRI兼容心脏起搏器的主动固定导线,第5天行1.5 T颈椎和胸部MRI并持续心电监护,基线和各序列后及成像后48小时检查电池阻抗、导线阻抗、自身振幅和起搏阈值,成像后4周对导线尖端周围心肌组织行组织病理学检查。结果显示装置位置和配置无显著变化,临床结局无异常,组织病理学见导线尖端周围心肌轻度炎症和坏死改变。作者结论认为,就装置配置与稳定性、临床结局及心肌组织病理学所见而言,植入的非MRI兼容起搏器行1.5 T MRI是安全的。

要点

  • 植入非MRI兼容起搏器的绵羊行1.5 T颈椎和胸部MRI,装置位置和配置无显著变化。
  • 两只绵羊临床结局无异常,成像后48小时装置参数检查未见异常报告。
  • 成像后4周组织病理学见导线尖端周围心肌轻度炎症和坏死改变。

关键影像表现/征象

  • 获得的序列包括T1加权、T2加权、T2梯度回波和扩散加权序列。
  • 一只绵羊采用OVO、VOO和VVI模式,另一只采用OAO、AOO和AAI模式,起搏和感知均为单极和双极配置。
  • MRI期间和之后未观察到装置位置或配置的显著变化。
  • 组织病理学检查见导线尖端周围心肌轻度炎症和坏死改变。

诊断标准与关键数值

  • 基线及各序列后、成像后48小时检查电池阻抗、起搏导线阻抗、自身振幅和起搏阈值。
  • 成像后4周对导线尖端周围心肌组织行组织病理学检查。

临床应用与阅片要点

  • 作者结论认为,1.5 T MRI对植入非MRI兼容起搏器的绵羊在装置配置与稳定性、临床结局和心肌组织病理学所见方面是安全的。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

AIM: To examine the clinical and histopathological consequences of MRI in sheep implanted with non-MRI-conditional cardiac pacemakers.

MATERIALS AND METHODS: Under general anesthesia, active fixation leads of two dual-chamber, non-MRI-conditional cardiac pacemakers (St. Jude Medical and Medtronic) were implanted either at the right ventricular apex or at the right atrium of two male sheep and connected to the V and A channels of the pacemakers, respectively. The generators were placed in cervical subcutaneous pockets. On day 5, both sheep underwent 1.5 T cervical and chest MRI with continuous electrocardiogram monitoring. Obtained sequences were T1-weighted (T1W), T2-weighted (T2W), T2-gradient echo and diffusion weighted (DW). The employed modes were OVO, VOO and VVI for one sheep and OAO, AOO and AAI for the other (unipolar and bipolar configuration of pacing and sensing for both). Battery impedance, pacing lead impedance, intrinsic amplitude and capture thresholds were checked at baseline and after each sequence, as well as 48 h after imaging. Histopathological examination of the cardiac tissue around the lead tip was performed 4 weeks post-imaging.

RESULTS: No significant changes in device position or configuration were observed during or after MRI. Clinical outcome was uneventful in both sheep. Minor inflammatory and necrotic changes were reported after histopathological examination of the cardiac tissue around the lead tip.

CONCLUSION: 1.5 T MRI of two implanted non-MRI-conditional pacemakers was found safe in terms of device configuration and stability, clinical outcome and cardiac tissue histopathological findings.